Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer.
Phys Med
; 117: 103203, 2024 Jan.
Article
en En
| MEDLINE
| ID: mdl-38171219
ABSTRACT
Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
Palabras clave
Texto completo:
1
Bases de datos:
MEDLINE
Asunto principal:
Neoplasias de la Mama
/
Radioterapia de Intensidad Modulada
/
Neoplasias de Mama Unilaterales
Tipo de estudio:
Etiology_studies
/
Observational_studies
/
Risk_factors_studies
Límite:
Female
/
Humans
Idioma:
En
Revista:
Phys Med
Asunto de la revista:
BIOFISICA
/
BIOLOGIA
/
MEDICINA
Año:
2024
Tipo del documento:
Article