Your browser doesn't support javascript.
loading
Comparison of depth of electrode insertion between cochleostomy and round window approach: a cadaveric study.
Jangra, Anupriya; Das, K Nidhin; Sharma, Vidhu; Timmaraju, Sailaja; Khera, Pushpinder; Tiwari, Sarbesh; Soni, Kapil; Choudhury, Bikram; Ghatak, Surajit; Dixit, Shilpi Gupta; Nayyar, Ashish Kumar; Goyal, Amit.
Afiliación
  • Jangra A; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India.
  • Das KN; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India.
  • Sharma V; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India.
  • Timmaraju S; Education and Training, MED-EL India Pvt Ltd, New Delhi, India.
  • Khera P; Department of Diagnostic and Interventional, Radiology, All India Institute of Medical Sciences, Jodhpur, India.
  • Tiwari S; Department of Diagnostic and Interventional, Radiology, All India Institute of Medical Sciences, Jodhpur, India.
  • Soni K; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India.
  • Choudhury B; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India.
  • Ghatak S; Department of Anatomy, All India Institute of Medical Sciences, Jodhpur, India.
  • Dixit SG; Department of Anatomy, All India Institute of Medical Sciences, Jodhpur, India.
  • Nayyar AK; Department of Anatomy, All India Institute of Medical Sciences, Jodhpur, India.
  • Goyal A; Department of Otorhinolaryngology, All India Institute of Medical Sciences, Jodhpur, India. meetugoyal@yahoo.com.
Eur Arch Otorhinolaryngol ; 281(7): 3547-3555, 2024 Jul.
Article en En | MEDLINE | ID: mdl-38294508
ABSTRACT

INTRODUCTION:

Round window approach and cochleostomy approach can have different depth of electrode insertion during cochlear implantation which itself can alter the audiological outcomes in cochlear implant.

OBJECTIVE:

The current study was conducted to determine the difference in the depth of electrode insertion via cochleostomy and round widow approach when done serially in same temporal bone.

METHODOLOGY:

This is a cross-sectional study conducted in the Department of Otorhinolaryngology in conjunction with Department of Anatomy and Department of Diagnostic and Interventional Radiology over a period of 1 year. 12-electrode array insertion was performed via either approach (cochleostomy or round window) in the cadaveric temporal bone. HRCT temporal bone scan of the implanted temporal bone was done and depth of insertion and various cochlear parameters were calculated.

RESULT:

A total of 12 temporal bones were included for imaging analysis. The mean cochlear duct length was 32.892 mm; the alpha and beta angles were 58.175° and 8.350°, respectively. The mean angular depth of electrode insertion via round window was found to be 325.2° (SD = 150.5842) and via cochleostomy 327.350 (SD = 112.79) degree and the mean linear depth of electrode insertion via round window was found to be 18.80 (SD = 4.4962) mm via cochleostomy 19.650 (SD = 3.8087) mm, which was calculated using OTOPLAN 1.5.0 software. There was a statically significant difference in linear depth of insertion between round window and cochleostomy. Although the angular depth of insertion was higher in CS group, there was no statistically significant difference with round window type of insertion.

CONCLUSION:

The depth of electrode insertion is one of the parameters that influences the hearing outcome. Linear depth of electrode insertion was found to be more in case of cochleostomy compared to round window approach (p = 0.075) and difference in case of angular depth of electrode insertion existed but not significant (p = 0.529).
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Ventana Redonda / Hueso Temporal / Cadáver / Implantes Cocleares / Cóclea / Implantación Coclear Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Eur Arch Otorhinolaryngol Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: India

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Ventana Redonda / Hueso Temporal / Cadáver / Implantes Cocleares / Cóclea / Implantación Coclear Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Eur Arch Otorhinolaryngol Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: India