Analytical Validation of a Volumetric Absorptive Microsampling Method for Therapeutic Drug Monitoring of the Oral Targeted Anticancer Agents, Abiraterone, Alectinib, Cabozantinib, Imatinib, Olaparib, and Sunitinib, and Metabolites.
Meertens, Marinda; de Vries, Niels; Rosing, Hilde; Steeghs, Neeltje; Beijnen, Jos H; Huitema, Alwin D R.
Afiliación
  • Meertens M; Department of Pharmacy & Pharmacology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, the Netherlands.
  • de Vries N; Department of Pharmacy & Pharmacology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, the Netherlands.
  • Rosing H; Department of Pharmacy & Pharmacology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, the Netherlands.
  • Steeghs N; Department of Medical Oncology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, the Netherlands.
  • Beijnen JH; Department of Pharmacy & Pharmacology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, the Netherlands.
  • Huitema ADR; Division of Pharmacoepidemiology and Clinical Pharmacology, Utrecht Institute for Pharmaceutical Sciences, Utrecht University, Utrecht, the Netherlands.
Ther Drug Monit ; 46(4): 494-502, 2024 Aug 01.
Article en En | MEDLINE | ID: mdl-38321598
ABSTRACT

BACKGROUND:

Volumetric Absorptive Microsampling (VAMS) is a useful tool for therapeutic drug monitoring (TDM) of oral targeted anticancer agents. VAMS aims to improve safety and efficacy by enabling at-home blood sample collection by patients. This study aimed to develop and validate an ultra-high performance liquid chromatography-tandem mass spectrometry method for the quantitative determination of abiraterone, alectinib, cabozantinib, imatinib, olaparib, sunitinib, and the metabolites, Δ(4)-abiraterone (D4A), alectinib-M4, imatinib-M1, and N -desethyl sunitinib, in dried whole blood samples using VAMS to support TDM.

METHODS:

After the collection of 10 µL of whole blood sample using the VAMS device, the analytes were extracted from the tip using methanol with shaking, evaporated, and reconstituted in acetonitrile0.1 mol/L ammonium hydroxide in water (11, vol/vol). The extracts were then analyzed using ultra-high performance liquid chromatography-tandem mass spectrometry. Validation experiments based on the ICH M10 guideline were carried out, and stability was evaluated under shipping and storage conditions. VAMS specimens were collected in the outpatient clinic to demonstrate the applicability of the assay.

RESULTS:

The validated range of the method was considered accurate and precise for all analytes. Accordingly, the validation experiments met the relevant requirements, except for cross-analyte interference. Based on the stability data, shipment can be performed at room temperature within 14 days after sample collection and the VAMS specimen can be stored up to 9 months at -20 and -70°C. Samples from 59 patients were collected at the hospital.

CONCLUSIONS:

The developed method could be used to successfully quantify the concentrations of abiraterone, D4A, alectinib, alectinib-M4, cabozantinib, imatinib, imatinib-M1, olaparib, sunitinib, and N -desethyl sunitinib within the validated range using VAMS. Therefore, the method can be used to estimate the dried whole blood-to-plasma ratios for TDM in the clinic.