Influence of limited mouth opening in children on intraoral scanning accuracy: An in vitro study.
Int J Paediatr Dent
; 34(6): 755-763, 2024 Nov.
Article
en En
| MEDLINE
| ID: mdl-38480519
ABSTRACT
BACKGROUND:
Although intraoral scanning is highly reliable, little is known about its accuracy in young children with limited mouth-opening ability.AIM:
To determine the accuracy of intraoral scans based on the degree of mouth opening.DESIGN:
To simulate mouth opening in children with primary dentition, three groups (n = 5 per group) were allocated by maximum mouth opening of 30, 37 and 40 mm. After the primary dentition model was connected to a dental phantom, intraoral scanning was performed using iTero and TRIOS4. The scanned files were digitally evaluated. Root mean square values were calculated to assess trueness and precision.RESULTS:
iTero showed deviations of three-dimensional trueness of 0.067 ± 0.008, 0.063 ± 0.001 and 0.065 ± 0.005 mm, and TRIOS4 of 0.07 ± 0.002, 0.064 ± 0.003 and 0.066 ± 0.002 mm in the 30, 37 and 40 mm groups, respectively. There were no significant differences in either mouth opening (p > .017) or the intraoral scanners (p > .05). The same statistical results were obtained for precision, with the exception of the 30 mm of mouth opening.CONCLUSIONS:
Within the limits of this study, limited mouth opening hardly influenced the accuracy of intraoral scanning.Palabras clave