Validation of a new extraction-free real-time PCR test to detect MPOX virus.
Diagn Microbiol Infect Dis
; 109(2): 116265, 2024 Jun.
Article
en En
| MEDLINE
| ID: mdl-38520832
ABSTRACT
The monkeypox (Mpox) virus has raised significant concerns given its recent spread with an increasing number of confirmed cases worldwide. In this study, we evaluated the performance of a laboratory developed test (LDT) using BioGX Xfree hMPXV/OPXV reagents for the qualitative detection of non-variola Orthopoxviruses and Mpox virus DNA, in swabs from human pustular or vesicular rash specimens. Analytical and clinical testing analysis were carried out on two different platforms the BD MAX™ System (BD Diagnostics) and the new pixl.16 Real-Time PCR Platform (BioGX), using a synthetic Mpox virus DNA (ATCC VR-3270SD) and residual clinical samples previously identified with an EUA approved Mpox real-time PCR assay. In the end, the Xfree hMPXV/OPXV LDT proved to be a sensitive, specific, and reproducible test for the detection of Mpox on both platforms evaluated with the pixl.16 having an advantage of a small footprint and providing faster TAT facilitated by an extraction-free workflow.