A setup for hard x-ray time-resolved resonant inelastic x-ray scattering at SwissFEL.
Struct Dyn
; 11(2): 024308, 2024 Mar.
Article
en En
| MEDLINE
| ID: mdl-38586277
ABSTRACT
We present a new setup for resonant inelastic hard x-ray scattering at the Bernina beamline of SwissFEL with energy, momentum, and temporal resolution. The compact R = 0.5 m Johann-type spectrometer can be equipped with up to three crystal analyzers and allows efficient collection of RIXS spectra. Optical pumping for time-resolved studies can be realized with a broad span of optical wavelengths. We demonstrate the performance of the setup at an overall â¼180 meV resolution in a study of ground-state and photoexcited (at 400 nm) honeycomb 5d iridate α-Li2IrO3. Steady-state RIXS spectra at the iridium L3-edge (11.214 keV) have been collected and are in very good agreement with data collected at synchrotrons. The time-resolved RIXS transients exhibit changes in the energy loss region <2 eV, whose features mostly result from the hopping nature of 5d electrons in the honeycomb lattice. These changes are ascribed to modulations of the Ir-to-Ir inter-site transition scattering efficiency, which we associate to a transient screening of the on-site Coulomb interaction.
Texto completo:
1
Bases de datos:
MEDLINE
Idioma:
En
Revista:
Struct Dyn
Año:
2024
Tipo del documento:
Article
País de afiliación:
Suiza