Comparison of peripheral blood automated hematopoietic progenitor cell count and flow cytometric CD34+ cell count.
Hagenkötter, Mischa; Mika, Thomas; Ladigan-Badura, Swetlana; Schork, Karin; Eisenacher, Martin; Schroers, Roland; Baraniskin, Alexander.
Afiliación
  • Hagenkötter M; Department of Medicine, Hematology and Oncology, UK Knappschaftskrankenhaus Bochum GmbH, Ruhr University of Bochum, Bochum, Germany.
  • Mika T; Department of Medicine, Hematology and Oncology, UK Knappschaftskrankenhaus Bochum GmbH, Ruhr University of Bochum, Bochum, Germany.
  • Ladigan-Badura S; Department of Medicine, Hematology and Oncology, UK Knappschaftskrankenhaus Bochum GmbH, Ruhr University of Bochum, Bochum, Germany.
  • Schork K; Department of Hematology and Oncology, Evangelisches Krankenhaus Hamm, Hamm, Germany.
  • Eisenacher M; Medizinisches Proteom-Center, Ruhr University Bochum, Bochum, Germany.
  • Schroers R; Medical Proteome Analysis, Center for Protein Diagnostics (PRODI), Ruhr-University Bochum, Bochum, Germany.
  • Baraniskin A; Medizinisches Proteom-Center, Ruhr University Bochum, Bochum, Germany.
J Clin Apher ; 39(3): e22114, 2024 Jun.
Article en En | MEDLINE | ID: mdl-38708583
ABSTRACT

BACKGROUND:

Stem cell apheresis in the context of autologous stem cell transplantation requires an accurate cluster of differentiantion 34 (CD34+) count determined by flow cytometry as the current gold standard. Since flow cytometry is a personnel and time-intensive diagnostic tool, automated stem cell enumeration may provide a promising alternative. Hence, this study aimed to compare automated hematopoietic progenitor enumeration carried out on a Sysmex XN-20 module compared with conventional flow cytometric measurements.

METHODS:

One hundred forty-three blood samples from 41 patients were included in this study. Correlation between the two methods was calculated over all samples, depending on leukocyte count and diagnosis.

RESULTS:

Overall, we found a high degree of correlation (r = 0.884). Furthermore, correlation was not impaired by elevated leukocyte counts (>10 000/µL, r = 0.860 vs <10 000/µL, r = 0.849; >20 000/µL, r = 0.843 vs <20 000/µL, r = 0.875). However, correlation was significantly impaired in patients with multiple myeloma (multiple myeloma r = 0.840 vs nonmyeloma r = 0.934).

SUMMARY:

Stem cell measurement carried out on the Sysmex XN-20 module provides a significant correlation with flow cytometry and might be implemented in clinical practice. In clinical decision-making, there was discrepancy of under 15% of cases. In multiple myeloma patients, XN-20 should be used with caution.