Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard.
Brambilla, Marco; Secco, Chiara; Dominietto, Marco; Matheoud, Roberta; Sacchetti, Gianmauro; Inglese, Eugenio.
Afiliação
  • Brambilla M; Health Physics Department, Azienda Ospedaliera Maggiore della Carità, Novara, Italy. marco.brambilla@maggioreosp.novara.it
J Nucl Med ; 46(12): 2083-91, 2005 Dec.
Article em En | MEDLINE | ID: mdl-16330574
ABSTRACT
UNLABELLED This article reports the results of performance measurements obtained for the lutetium oxyorthosilicate (LSO)-based whole-body PET/CT scanner Biograph 16 HI-REZ with the National Electrical Manufacturers Association (NEMA) NU 2-2001 standard. The Biograph 16 HI-REZ combines a multislice (16-slice) spiral CT scanner with a PET scanner composed of 24.336 LSO crystals arranged in 39 rings. The crystal dimensions are 4.0x4.0x20 mm3, and the crystals are organized in 13x13 blocks coupled to 4 photomultiplier tubes each. The 39 rings allow the acquisition of 81 images 2.0 mm thick, covering an axial field of view of 162 mm. The low- and high-energy thresholds are set to 425 and 650 keV, respectively, acquiring data within a 4.5-ns-wide coincidence window.

METHODS:

Performance measurements for the LSO-based PET/CT scanner were obtained with the NEMA NU 2-2001 standard, taking into account issues deriving from the presence of intrinsic radiation.

RESULTS:

The results obtained with the NEMA NU 2-2001 standard measurements were as follows average transverse and axial spatial resolutions (full width at half maximum) at 1 cm and at 10 cm off axis of 4.61 (5.10) mm and 5.34 (5.91) mm, respectively; average sensitivity of 4.92 counts per second per kilobecquerel for the 2 radial positions (0 and 10 cm); 34.1% system scatter fraction; and peak noise equivalent count (NEC) rates of 84.77 kilocounts per second (kcps) at 28.73 kBq/mL (k=1 in the NEC formula; noiseless random correction) and 58.71 kcps at 21.62 kBq/mL (k=2; noisy random correction).

CONCLUSION:

The new integrated PET/CT system Biograph 16 HI-REZ has good overall performance, with, in particular, a high resolution, a low scatter fraction, and a very good NEC response.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomógrafos Computadorizados / Silicatos / Imageamento Tridimensional / Tomografia por Emissão de Pósitrons / Lutécio Tipo de estudo: Diagnostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Nucl Med Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomógrafos Computadorizados / Silicatos / Imageamento Tridimensional / Tomografia por Emissão de Pósitrons / Lutécio Tipo de estudo: Diagnostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Nucl Med Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Itália