Analysis of the nonclinical telemetered ECG: impact of logging rate and RR bin width in the dog and cynomolgus monkey.
Holzgrefe, Henry H; Cavero, Icilio; Gleason, Carol R.
Afiliação
  • Holzgrefe HH; Bristol-Myers Squibb Pharmaceutical Research Institute, United States. Henry.Holzgrefe@Roche.com
J Pharmacol Toxicol Methods ; 56(1): 34-42, 2007.
Article em En | MEDLINE | ID: mdl-17258913
ABSTRACT

INTRODUCTION:

Species-dependent ECG differences may affect QT interval analysis. Among these are the range of QT and RR values, heart rate variability, and respiratory sinus arrhythmia (SA). Importantly, the effects of data logging rates and RR bin width (BW) on QT analysis have not been characterized in dogs and nonhuman primates.

METHODS:

Digital ECGs were collected for 18-21 h in telemetered dogs (n=7) and cynomolgus monkeys (n=7) employing epicardial ECG leads and analyzed by computerized algorithms. ECG intervals were determined employing beat-to-beat (1 ms BW) and 10 s logging rates (for 1, 10, 20, and 50 ms BW). Diurnal heart rate variability was defined as the beat-to-beat RR ratio (VR) where SA was defined as >+/-10% change in VR.

RESULTS:

Canine beat-to-beat QT-RR data were curvilinear and plateaued, with multiple RR values associated to any given QT for RR>or=900 ms. Cynomolgus QT-RR data collected as beat-to-beat or at a 10 s logging rate were linear for all RR intervals. During the day, SA comprised 62.9+/-2.4% and 11.4+/-6.1% of total beats in the dog and cynomolgus, respectively, with no diurnal/nocturnal differences in either species. QT interval changes varied with BW such that 5 ms resolution was maintained for BW dog and cynomolgus, respectively.

DISCUSSION:

Differences in the QT interval, primarily due to SA, were responsible for skewed canine beat-to-beat QT-RR relationships. A 10 s logging rate eliminated this confounding factor, enabling the derivation of accurate QT rate-corrections. Thus, the uniform application of a 10 s logging rate in conjunction with a 10 ms RR BW ensured consistent and accurate QT analysis in both species. In conclusion, optimal BWs were defined which maintained beat-to-beat accuracy, while maximizing the number of QT-RR observations incorporated into QT-RR rate-correction models, a critical factor for robust probabilistic QT rate-corrections.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Eletrocardiografia / Frequência Cardíaca Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Pharmacol Toxicol Methods Assunto da revista: FARMACOLOGIA / TOXICOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Eletrocardiografia / Frequência Cardíaca Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Pharmacol Toxicol Methods Assunto da revista: FARMACOLOGIA / TOXICOLOGIA Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Estados Unidos