Scattered neutron dose equivalent from an active scanning proton beam delivery system.
Hecksel, Draik; Sandison, George A; Farr, Jonathan B; Edwards, Andrew C.
Afiliação
  • Hecksel D; School of Health Sciences, Purdue University, West Lafayette, Indiana, USA.
Australas Phys Eng Sci Med ; 30(4): 326-30, 2007 Dec.
Article em En | MEDLINE | ID: mdl-18274074
A study of neutron production from a novel active scanning proton beam delivery system at the Midwest Proton Radiotherapy Institute (MPRI) has been performed. The neutron dose equivalent was determined using a neutron rem (roentgen equivalent in man) detector which has an upper energy limit of 10 MeV. Measurement were taken at 0, 45, and 90 degrees from the proton beam central axis and for various proton beam energies (127-208 MeV) and scanned field sizes (25-144 cm2). The maximum neutron dose observed was 0.43 mSv / (proton treatment Gy) at 90 degrees from the beam axis for a beam energy of 208.4 MeV and a scanned field size of 144 cm2. It is still possible to further mitigate this secondary neutron dose during treatment by optimizing parameters within the treatment nozzle and using shielding.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Terapia com Prótons / Nêutrons Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google