Your browser doesn't support javascript.
loading
Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity.
Horváth, Iván G; Németh, Adám; Lenkey, Zsófia; Alessandri, Nicola; Tufano, Fabrizio; Kis, Pál; Gaszner, Balázs; Cziráki, Attila.
Afiliação
  • Horváth IG; aHeart Institute, Medical School, University of Pécs, Pécs, Hungary bDepartment of Cardiology, University of Rome La Sapienza, Polo Pontino, Italy.
J Hypertens ; 28(10): 2068-75, 2010 Oct.
Article em En | MEDLINE | ID: mdl-20651604
BACKGROUND: The importance of measuring aortic pulse wave velocity (PWVao), aortic augmentation index (Aix) and central systolic blood pressure (SBPao) has been shown under different clinical conditions; however, information on these parameters is hard to obtain. The aim of this study was to evaluate the accuracy of a new, easily applicable oscillometric device (Arteriograph), determining these parameters simultaneously, against invasive measurements. METHODS: Aortic Aix, SBPao and PWVao were measured invasively during cardiac catheterization in 16, 55 and 22 cases, respectively, and compared with the values measured by the Arteriograph. RESULTS: We found strong correlation between the invasively measured aortic Aix and the oscillometrically measured brachial Aix on either beat-to-beat or mean value per patient basis (r = 0.9, P < 0.001; r = 0.94, P < 0.001), which allowed the noninvasive calculation of the aortic Aix without using generalized transfer function. Similarly strong correlation (r = 0.95, P < 0.001) was found between the invasively measured and the noninvasively calculated central SBPao; furthermore, the BHS assessment of the paired differences fulfilled the 'B' grading. The PWVao values measured invasively and by Arteriograph were 9.41 ± 1.8 m/s and 9.46 ± 1.8 m/s, respectively (mean ± SD); furthermore, the Pearson's correlation was 0.91 (P < 0.001). The limits of agreement were 11.4% for aortic Aix and 1.59 m/s for PWVao. CONCLUSION: Aix, SBPao and PWVao, measured oscillometrically, showed strong correlation with the invasively obtained values. The observed limits of agreement are encouragingly low for accepting the method for clinical use. Our results suggest that the PWVao values, measured by Arteriograph, are close to the true aortic PWV, determined invasively.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Aorta / Oscilometria / Fluxo Sanguíneo Regional / Velocidade do Fluxo Sanguíneo / Pressão Sanguínea / Angiografia Tipo de estudo: Clinical_trials Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: J Hypertens Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Aorta / Oscilometria / Fluxo Sanguíneo Regional / Velocidade do Fluxo Sanguíneo / Pressão Sanguínea / Angiografia Tipo de estudo: Clinical_trials Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: J Hypertens Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Itália