Your browser doesn't support javascript.
loading
Asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead detected by home monitoring system.
Migliore, Federico; Leoni, Loira; Torregrossa, Gianluca; Guglielmi, Cosimo; Tarantini, Giuseppe; Buja, Gianfranco; Iliceto, Sabino; Corrado, Domenico.
Afiliação
  • Migliore F; Department of Cardiac, Thoracic and Vascular Sciences, University of Padua, Padova, Italy.
J Electrocardiol ; 43(6): 673-5, 2010.
Article em En | MEDLINE | ID: mdl-20888001
Right ventricular lead perforation is a rare but serious and potentially life-threatening complication of pacemaker or defibrillator lead implantation. This report describes a patient with Brugada syndrome in whom the diagnosis of asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead was detected 12 days after implantation, thanks to a report from home monitoring system. The patient was admitted to our institution, where the lead was explanted and replaced. This case illustrates the potential lifesaving benefit of the home monitoring system in patients with implantable cardioverter defibrillator.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Ferimentos Penetrantes / Desfibriladores Implantáveis / Eletrodos Implantados / Traumatismos Cardíacos / Ventrículos do Coração Limite: Humans / Male / Middle aged Idioma: En Revista: J Electrocardiol Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Ferimentos Penetrantes / Desfibriladores Implantáveis / Eletrodos Implantados / Traumatismos Cardíacos / Ventrículos do Coração Limite: Humans / Male / Middle aged Idioma: En Revista: J Electrocardiol Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Itália