Direct microscopic observation of striations in a fractured section of a sirolimus-eluting stent (Cypher Bx Velocity®) indicates induction of stent fracture by continuous shear stress.
Endo, Akira; Ishikawa, Tetsuya; Suzuki, Teruhiko; Kashiwagi, Yusuke; Mutoh, Makoto.
Afiliação
  • Endo A; Department of Cardiology, Saitama Prefecture Cardiovascular Respiratory Center, 1696 Itai, Kumagaya, Saitama, Japan.
Int Heart J ; 52(4): 248-51, 2011.
Article em En | MEDLINE | ID: mdl-21828954
ABSTRACT
A 73-year-old woman with severe congestive heart failure was treated by implantation with a sirolimus-eluting stent (SES; Cypher Bx Velocity(®)) in the left main coronary artery (LMCA) using the staged T-stent, kissing balloon, and hugging balloon techniques. Follow-up coronary multislice computed tomography after 10 months revealed that SES was completely fractured in 2 directions; the fractured stent appeared in the shape of the letter "L" and had migrated into the aorta. An SES fragment was surgically removed and subsequent electron microscopy revealed striations (striped patterns in fractured sections) on the fracture plane, indicating continuous shear stress after SES implantation in the LMCA. This case provides direct evidence of continuous shear stress on the SES and indicates the necessity of improving the structure of the stent such that it can withstand shear stress.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Falha de Prótese / Microscopia Eletrônica / Angioplastia Coronária com Balão / Migração de Corpo Estranho / Sirolimo / Stents Farmacológicos / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Revista: Int Heart J Assunto da revista: CARDIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Falha de Prótese / Microscopia Eletrônica / Angioplastia Coronária com Balão / Migração de Corpo Estranho / Sirolimo / Stents Farmacológicos / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Revista: Int Heart J Assunto da revista: CARDIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Japão