Your browser doesn't support javascript.
loading
QRS subtraction for atrial electrograms: flat, linear and spline interpolations.
Ahmad, A; Salinet, J L; Brown, P; Tuan, J H; Stafford, P; Ng, G Andre; Schlindwein, F S.
Afiliação
  • Ahmad A; Department of Engineering, University of Leicester, Leicester LE17RH, UK. aa384@le.ac.uk
Med Biol Eng Comput ; 49(11): 1321-8, 2011 Nov.
Article em En | MEDLINE | ID: mdl-21959592
ABSTRACT
The main objective of this article is to implement and compare QRS subtraction techniques for intra-cardiac atrial electrograms based on using the surface ECG as a reference. A band-pass filter between 8 and 20 Hz followed by rectification, and then a low-pass filter at 6 Hz are used for QRS detection. QRS subtraction was performed using three different approaches flat, linear and spline interpolations. QRS subtraction affects the power of the signals but it normally does not affect the dominant frequency. The average power of the atrial electrograms after QRS subtraction is significantly reduced for frequencies above 10 Hz.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Fibrilação Atrial / Processamento de Sinais Assistido por Computador / Técnicas Eletrofisiológicas Cardíacas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Med Biol Eng Comput Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Fibrilação Atrial / Processamento de Sinais Assistido por Computador / Técnicas Eletrofisiológicas Cardíacas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Med Biol Eng Comput Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Reino Unido