Spatiotemporal image correlation artifacts in an in vitro model.
Haak, Monique C; Uittenbogaard, Lukas B; van Vugt, John M G.
Afiliação
  • Haak MC; Departmentof Obstetrics and Gynecology, VU University Medical Center, Amsterdam, the Netherlands. m.c.haak@lumc.nl
J Ultrasound Med ; 30(10): 1411-4, 2011 Oct.
Article em En | MEDLINE | ID: mdl-21968493
An experimental in vitro setting with a latex miniature balloon was designed to test the accuracy of volumetric measurements by spatiotemporal image correlation. Two-dimensional images clearly showed the round balloon as a thin echogenic ring in a translucent area. Four-dimensional reconstructed images, however, showed a severely distorted balloon. The artifacts disappeared when the surroundings of the balloons were made echogenic, mimicking the in vivo setting. We hypothesize that the artifacts were the result of gating errors. These experiments can be relevant for analysis of spatiotemporal image correlation volumes in daily practice.
Assuntos
Buscar no Google
Bases de dados: MEDLINE Assunto principal: Ultrassonografia / Artefatos Tipo de estudo: Diagnostic_studies Idioma: En Revista: J Ultrasound Med Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Holanda
Buscar no Google