The locking attachment plate for proximal fixation of periprosthetic femur fractures--a biomechanical comparison of two techniques.
Lenz, Mark; Windolf, Markus; Mückley, Thomas; Hofmann, Gunther O; Wagner, Michael; Richards, Robert G; Schwieger, Karsten; Gueorguiev, Boyko.
Afiliação
  • Lenz M; AO Research Institute Davos, Clavadelerstrasse 8, 7270 Davos-Platz, Switzerland. mark.lenz@aofoundation.org
Int Orthop ; 36(9): 1915-21, 2012 Sep.
Article em En | MEDLINE | ID: mdl-22638607
ABSTRACT

PURPOSE:

Mechanical properties of a locking attachment plate construct (LAP-LCP), allowing bicortical screw placement laterally to the prosthesis stem, are compared to a cerclage-LCP construct.

METHODS:

Eight right synthetic femora with implanted uncemented hip endoprosthesis were cut distally and fixed with LCP, monocortical locking screws and either LAP (n = 4) or cerclage (n = 4). Cyclic testing was performed with monotonically increasing sinusoidal load until failure. Relative movements at the plate-femur interface were registered by motion tracking. Statistical differences were detected by unpaired t-test and general linear model repeated measures.

RESULTS:

Stiffness of the LAP-LCP was significantly higher at the beginning (875.4 N/mm ± 29.8) and after 5000 cycles (1213.0 N/mm ± 101.1) compared to the cerclage-LCP (644.96 N/mm ± 50.1 and 851.9 N/mm ± 81.9), with p = 0.013. Relative movements for AP-bending (B) and axial translation (T) of the LAP-LCP at the beginning (0.07° ± 0.02, 0.20 mm ± 0.08), after 500 cycles (0.16° ± 0.10, 0.26 mm ± 0.07) and after 5000 cycles (0.26° ± 0.11, 0.31 mm ± 0.07) differed significantly from the cerclage-LCP (beg. 0.26° ± 0.04, 0.28 mm ± 0.05; 500 cyc 0.47° ± 0.03, 0.53 mm ± 0.07; 5000 cyc. 0.63° ± 0.18, 0.79 mm ± 0.13), with B p = 0.02, T p = 0.04. Relative movements for medial bending were not significantly different between the two constructs. Cycles to failure (criterion 1 mm axial translation) differed significantly between LAP-LCP (19,519 ± 1,758) and cerclage-LCP (11,265 ± 2,472), with p = 0.035.

CONCLUSIONS:

Biomechanically, the LAP-LCP construct improves proximal fixation of periprosthetic fractures compared to the cerclage-LCP construct.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Placas Ósseas / Teste de Materiais / Fios Ortopédicos / Fraturas Periprotéticas / Fraturas do Fêmur / Fixação Interna de Fraturas Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Int Orthop Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Placas Ósseas / Teste de Materiais / Fios Ortopédicos / Fraturas Periprotéticas / Fraturas do Fêmur / Fixação Interna de Fraturas Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Int Orthop Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Suíça