Your browser doesn't support javascript.
loading
Non-contrast-enhanced renal and abdominal MR angiography using velocity-selective inversion preparation.
Shin, Taehoon; Worters, Pauline W; Hu, Bob S; Nishimura, Dwight G.
Afiliação
  • Shin T; Department of Electrical Engineering, Stanford University, Stanford, California 94305, USA. taehoons@stanford.edu
Magn Reson Med ; 69(5): 1268-75, 2013 May.
Article em En | MEDLINE | ID: mdl-22711643
ABSTRACT
Non-contrast-enhanced MR angiography is a promising alternative to the established contrast-enhanced approach as it reduces patient discomfort and examination costs and avoids the risk of nephrogenic systemic fibrosis. Inflow-sensitive slab-selective inversion recovery imaging has been used with great promise, particularly for abdominal applications, but has limited craniocaudal coverage due to inflow time constraints. In this work, a new non-contrast-enhanced MR angiography method using velocity-selective inversion preparation is developed and applied to renal and abdominal angiography. Based on the excitation k-space formalism and Shinnar-Le-Roux transform, a velocity-selective excitation pulse is designed that inverts stationary tissues and venous blood while preserving inferiorly flowing arterial blood. As the magnetization of the arterial blood in the abdominal aorta and iliac arteries is well preserved during the magnetization preparation, artery visualization over a large abdominal field of view is achievable with an inversion delay time that is chosen for optimal background suppression. Healthy volunteer tests demonstrate that the proposed method significantly increases the extent of visible arteries compared with the slab-selective approach, covering renal arteries through iliac arteries over a craniocaudal field of view of 340 mm.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Aorta Abdominal / Artéria Renal / Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Angiografia por Ressonância Magnética / Artéria Ilíaca Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Aorta Abdominal / Artéria Renal / Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Angiografia por Ressonância Magnética / Artéria Ilíaca Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos