Your browser doesn't support javascript.
loading
Permanent His-bundle pacing maintains long-term ventricular synchrony and left ventricular performance, unlike conventional right ventricular apical pacing.
Catanzariti, Domenico; Maines, Massimiliano; Manica, Anna; Angheben, Carlo; Varbaro, Annamaria; Vergara, Giuseppe.
Afiliação
  • Catanzariti D; Division of Cardiology, Santa Maria del Carmine Hospital, Rovereto (TN), Italy.
Europace ; 15(4): 546-53, 2013 Apr.
Article em En | MEDLINE | ID: mdl-22997222
ABSTRACT

AIMS:

Right ventricular apical pacing (RVAP) may be deleterious, determining abnormal left ventricular (LV) electrical activation and progressive LV dysfunction. Permanent His-bundle pacing (HBP) has been proposed to prevent this detrimental effect. The aim of our study was to compare the long-term effects of HBP on LV synchrony and systolic performance with those of RVAP in the same group of patients.

METHODS:

Our analysis included 26 patients who received both an HBP lead and an RVAP lead, as backup, in our electrophysiology laboratory between 2004 and 2007. After implantation, all devices were programmed to obtain HBP. An intra-patient comparison of the effects of HBP and RVAP on LV dyssynchrony and function was performed at the last available follow-up examination.

RESULTS:

After a mean of 34.6 ± 11 months, the pacing modality was temporarily switched to RVAP. During RVAP, LV ejection fraction significantly decreased (50.1 ± 8.8% vs. 57.3 ± 8.5%, P < 0.001), mitral regurgitation significantly increased (22.5 ± 10.9% vs.16.3 ± 12.4%; P = 0.018), and inter-ventricular delay significantly worsened (33.4 ± 19.5 ms vs. 7.1 ± 4.7 ms, P = 0.003) in comparison with HBP. However, the myocardial performance index was not statistically different between the two pacing modalities (P = 0.779). No asynchrony was revealed by tissue Doppler imaging during HBP, while during RVAP the asynchrony index was significantly higher in both the four-chamber (125.8 ± 63.9 ms; P = 0.035 vs. HBP) and two-chamber (126 ± 86.5 ms; P = 0.037 vs. HBP) apical views.

CONCLUSION:

His-bundle pacing has long-term positive effects on inter- and intra-ventricular synchrony and ventricular contractile performance in comparison with RVAP. It prevents asynchronous pacing-induced LV ejection fraction depression and mitral regurgitation.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Fascículo Atrioventricular / Estimulação Cardíaca Artificial / Função Ventricular Esquerda / Função Ventricular Direita Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Fascículo Atrioventricular / Estimulação Cardíaca Artificial / Função Ventricular Esquerda / Função Ventricular Direita Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália