Preliminary results from BCG and ECG measurements in the heart failure clinic.
Giovangrandi, Laurent; Inan, Omer T; Banerjee, Dipanjan; Kovacs, Gregory T A.
Afiliação
  • Giovangrandi L; Electrical Engineering Department, Stanford University, Stanford, CA 94305, USA. giovan@stanford.edu
Article em En | MEDLINE | ID: mdl-23366751
We report on the preliminary deployment of a bathroom scale-based ballistocardiogram (BCG) system for the in-hospital monitoring of patients with heart failure. These early trials provided valuable insights into the challenges and opportunities for such monitoring. In particular, the need for robust algorithms and adapted BCG metric is suggested. The system was designed to be robust and user-friendly, with dual ballistocardiogram (BCG) and electrocardiogram (ECG) capabilities. The BCG was measured from a modified bathroom scale, while the ECG (used as timing reference) was measured using dry handlebar electrodes. The signal conditioning and digitization circuits were USB-powered, and data acquisition performed using a netbook. Four patients with a NYHA class III at admission were measured daily for the duration of their treatment at Stanford hospital. A measure of BCG quality, in essence a quantitative implementation of the BCG classes originally defined in the 1950s, is proposed as a practical parameter.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Balistocardiografia / Eletrocardiografia / Insuficiência Cardíaca Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Balistocardiografia / Eletrocardiografia / Insuficiência Cardíaca Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos