Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved?
Schindera, Sebastian T; Odedra, Devang; Raza, Syed Arsalan; Kim, Tae Kyoung; Jang, Hyun-Jung; Szucs-Farkas, Zsolt; Rogalla, Patrik.
Afiliação
  • Schindera ST; Department of Medical Imaging, University Health Network, University of Toronto, Toronto, Ont, Canada; Clinic of Radiology and Nuclear Medicine, University Hospital Basel, Petersgraben 4, CH-4031 Basel, Switzerland.
Radiology ; 269(2): 511-8, 2013 Nov.
Article em En | MEDLINE | ID: mdl-23788715
PURPOSE: To compare the low-contrast detectability and image quality of computed tomography (CT) at different radiation dose levels reconstructed with iterative reconstruction (IR) and filtered back projection (FBP). MATERIALS AND METHODS: A custom liver phantom with 12 simulated hypoattenuating tumors (diameters of 5, 10, 15, and 20 mm; tumor-to-liver contrast values of -10, -20, and -40 HU) was designed. The phantom was scanned with a standard abdominal CT protocol with a volume CT dose index of 21.6 mGy (equivalent 100% dose) and four low-dose protocols (20%, 40%, 60%, and 80% of the standard protocol dose). CT data sets were reconstructed with IR and FBP. Image noise was measured, and the tumors' contrast-to-noise ratios (CNRs) were calculated. Tumor detection was independently assessed by three radiologists who were blinded to the CT technique used. A total of 840 simulated tumors were presented to the radiologists. Statistical analyses included analysis of variance. RESULTS: IR yielded an image noise reduction of 43.9%-63.9% and a CNR increase of 74.1%-180% compared with FBP at the same dose level (P < .001). The overall sensitivity for tumor detection was 64.7%-85.3% for IR and 66.3%-85.7% for FBP at the 20%-100% doses, respectively. There was no significant difference in the sensitivity for tumor detection between IR and FBP at the same dose level (P = .99). The sensitivity of the protocol at the 20% dose with FBP and IR was significantly lower than that of the protocol at the 100% dose with FBP and IR (P = .019). CONCLUSION: As the radiation dose at CT decreases, the IR algorithm does not preserve the low-contrast detectability. SUPPLEMENTAL MATERIAL: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.13122349/-/DC1.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Doses de Radiação / Algoritmos / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Radiology Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Doses de Radiação / Algoritmos / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Radiology Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Suíça