Method for the calibration of the spectral irradiance of tungsten filament transfer standard sources traceable to synchrotron radiation.
Anevsky, Sergey; Krutikov, Vladimir; Minaeva, Olga; Minaev, Roman; Senin, Dmitriy; Hollandt, Jörg; Taubert, Dieter R.
Afiliação
  • Anevsky S; All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI), 119361 Ozernaya St. 46, Moscow, Russia.
Appl Opt ; 52(21): 5152-7, 2013 Jul 20.
Article em En | MEDLINE | ID: mdl-23872760
ABSTRACT
The spectral irradiance calibration of tungsten strip and spiral filament lamps applying synchrotron radiation revealed that the spectral irradiance in the wavelength range from 280 to 400 nm can be well approximated by blackbody radiation according to Planck's law. Consequently, the spectral irradiance of the filament lamp can then be described by an effective irradiance temperature, which would be beneficial for practical measurements. Including the emissivity of tungsten into the approximation, the model can be expanded to visible and near-infrared wavelength regions. The effective irradiance temperature dependence of the lamp current was investigated and appeared to be close to linear.

Texto completo: 1 Bases de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Revista: Appl Opt Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Federação Russa