Development of a method for the determination of cefovecin in plasma by HPLC.
Cox, Sherry; Martin, Ali; Hamill, Mallery; Pistole, Nealy; Steeil, James; Schumacher, Juergen.
Afiliação
  • Cox S; The University of Tennessee, College of Veterinary Medicine, Department of Biomedical and Diagnostic Sciences, 2407 River Drive, Knoxville, TN, 37996, USA.
Biomed Chromatogr ; 28(8): 1126-30, 2014 Aug.
Article em En | MEDLINE | ID: mdl-24420236
ABSTRACT
A simple high-performance liquid chromatography method for the determination of cefovecin in small volume plasma has been developed. Following solid-phase extraction using Oasis HLB cartridges, samples were separated by reverse-phase high-performance liquid chromatography on an XBridge C8 (3.5 µm) 4.6 × 250 mm column and quantified using ultraviolet detection at 280 nm. The mobile phase was a mixture of 10 mm ammonium acetate (pH 3.5) and acetonitrile (8911), with a flow rate of 0.85 mL/min. The standard curve ranged from 0.1 to 200 µg/mL. Intra- and Inter-assay variability for cefovecin was <10%, and the average recovery was >90%. The lower limit of quantitation was 0.1 µg/mL. This method was successfully applied to the analysis of cefovecin samples at our institution. This is also the first fully validated method with an internal standard that does not use mass spectrometry.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Cefalosporinas / Cromatografia Líquida de Alta Pressão Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Biomed Chromatogr Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Cefalosporinas / Cromatografia Líquida de Alta Pressão Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Biomed Chromatogr Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Estados Unidos