Lack of interference of electromagnetic navigation bronchoscopy to implanted cardioverter-defibrillator: in-vivo study.
Magnani, Andrea; Balbo, Piero; Facchini, Emanuela; Occhetta, Eraldo; Marino, Paolo.
Afiliação
  • Magnani A; Department of Cardiology, University Hospital 'Maggiore della Carità', Novara, Italy andrea.magnani@maggioreosp.novara.it.
  • Balbo P; Department of Pneumology, University Hospital 'Maggiore della Carità', Novara, Italy.
  • Facchini E; Department of Cardiology, University Hospital 'Maggiore della Carità', Novara, Italy.
  • Occhetta E; Department of Cardiology, University Hospital 'Maggiore della Carità', Novara, Italy.
  • Marino P; Department of Cardiology, University Hospital 'Maggiore della Carità', Novara, Italy.
Europace ; 16(12): 1767-71, 2014 Dec.
Article em En | MEDLINE | ID: mdl-25082949
AIMS: Electromagnetic navigation bronchoscopy (ENB) (Superdimension) is a diagnostic and therapeutic tool in patients with lung lesions. Very small data are available about potential interference of ENB magnetic field to implanted cardioverter-defibrillators (ICDs) and any documentation of ICD behaviour if a ventricular tachyarrhythmia occurs during ENB is lacking. We tested a number of selected ICDs to assess if any interference occurs by ENB magnetic field on detection of clinical ventricular fibrillation and shock delivery. METHODS AND RESULTS: Thirteen patients undergoing an ICD implantation or elective replacement with a clinical indication to assess the efficacy of defibrillation underwent: (i) real-time telemetric recording from ICD during ENB activation to detect possible noise; (ii) defibrillation test during exposure to ENB board-generated magnetic field. All tested ICDs showed no noise detection at maximum sensitivity level. Induced ventricular fibrillation was correctly detected and cured by implanted device. No change in programmed ICD parameters was induced by exposure to ENB magnetic field. CONCLUSION: All tested ICDs correctly operated and rescued the patients from induced ventricular fibrillation during ENB. Electromagnetic navigation bronchoscopy appears to be safe in heart patients with an ICD; however, close cardiac monitoring of these patients during ENB must be ensured as correct behaviour of all existing ICDs can only be presumed from compliance of the manufacturer to International Standards which establish procedures for electromagnetic interference checking on implantable devices on different ranges of frequency.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Broncoscopia / Artefatos / Desfibriladores Implantáveis / Campos Eletromagnéticos / Falha de Equipamento Tipo de estudo: Guideline Limite: Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Broncoscopia / Artefatos / Desfibriladores Implantáveis / Campos Eletromagnéticos / Falha de Equipamento Tipo de estudo: Guideline Limite: Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Itália