Evaluation of image lag in a flat-panel, detector-equipped cardiovascular X-ray machine using a newly developed dynamic phantom.
Sato, Hisaya; Kondo, Keisuke; Kato, Kyoichi; Nakazawa, Yasuo.
Afiliação
  • Sato H; 1)Graduate School of Nursing and Rehabilitation Sciences, Showa University 2)Graduate School of Medical Health Sciences, Komazawa University. hisaya@cmed.showa-u.ac.jp.
J Appl Clin Med Phys ; 16(2): 5213, 2015 Mar 08.
Article em En | MEDLINE | ID: mdl-26103192
ABSTRACT
We developed a dynamic phantom for use in routine checks. This phantom can be used to physically evaluate image lag that occurs in dynamic images. It has a unique measurement method. ROIs on the target are chosen, and, with the position of ROIs fixed on the image, changes in pixel value are detected physically when the target passes through the ROIs over time and perceived as image lag. Thus, it was possible to physically detect different intensities of image lag lasting less than one second while maintaining the same intensity trends. The checking technique we propose with the dynamic phantom that we developed could be effective for routine checking of fluoroscopy X-ray machines, and could become an established method.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Fluoroscopia / Interpretação de Imagem Radiográfica Assistida por Computador / Doenças Cardiovasculares / Imagens de Fantasmas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Fluoroscopia / Interpretação de Imagem Radiográfica Assistida por Computador / Doenças Cardiovasculares / Imagens de Fantasmas Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2015 Tipo de documento: Article