Improvement of temporal signal-to-noise ratio of GRAPPA accelerated echo planar imaging using a FLASH based calibration scan.
Talagala, S Lalith; Sarlls, Joelle E; Liu, Siyuan; Inati, Souheil J.
Afiliação
  • Talagala SL; NIH MRI Research Facility, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
  • Sarlls JE; NIH MRI Research Facility, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
  • Liu S; Language Section, Voice, Speech and Language Branch, National Institute on Deafness and Other Communication Disorders, National Institutes of Health, Bethesda, Maryland, USA.
  • Inati SJ; Functional MRI Facility, National Institute of Mental Health, National Institutes of Health, Bethesda, Maryland, USA.
Magn Reson Med ; 75(6): 2362-71, 2016 06.
Article em En | MEDLINE | ID: mdl-26192822
ABSTRACT

PURPOSE:

To demonstrate that the temporal signal-to-noise ratio (SNR) of generalized autocalibrating partially parallel acquisitions (GRAPPA) accelerated echo planar imaging (EPI) can be enhanced and made more spatially uniform by using a fast low angle shot (FLASH) based calibration scan.

METHODS:

EPI of a phantom and human brains were acquired at 3 Tesla without and with GRAPPA acceleration factor of 2. The GRAPPA accelerated data were reconstructed using calibration scans acquired with EPI and FLASH acquisition schemes. The increase in temporal signal fluctuation due to GRAPPA reconstruction was quantified and compared. Simulated g-factor maps were also created for different calibration scans.

RESULTS:

GRAPPA accelerated phantom data exhibited areas with high g values when using the EPI based calibration for reconstruction. The g-factor maps were uniform when using the FLASH calibration scan. g was greater than 1.1 in 74% of pixels in 64 × 64 data reconstructed with the EPI calibration compared with only 15% when using the FLASH calibration scan. Human data also showed abnormally high g regions when using the EPI calibration but not when using the FLASH calibration scan. Use of the FLASH calibration scan increased the whole brain temporal SNR by ∼12% without affecting the image quality. Experimental observations were confirmed by simulations.

CONCLUSION:

A calibration scan based on a FLASH acquisition scheme can be used to improve the temporal SNR of GRAPPA accelerated EPI time series. Magn Reson Med 752362-2371, 2016. Published 2015. This article is a U.S. Government work and is in the public domain in the USA.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imagem Ecoplanar Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imagem Ecoplanar Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos