Multilayer based soft-x-ray polarimeter at MAX IV Laboratory.
Grizolli, Walan; Laksman, Joakim; Hennies, Franz; Jensen, Brian Norsk; Nyholm, Ralf; Sankari, Rami.
Afiliação
  • Grizolli W; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
  • Laksman J; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
  • Hennies F; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
  • Jensen BN; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
  • Nyholm R; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
  • Sankari R; MAX IV Laboratory, P.O. Box 118, SE-22100 Lund, Sweden.
Rev Sci Instrum ; 87(2): 025102, 2016 Feb.
Article em En | MEDLINE | ID: mdl-26931886
ABSTRACT
A high precision five rotation-axes polarimeter using transmission multilayers as polarizers and reflection multilayers as analyzers has been designed and manufactured. To cover the extreme ultraviolet regime, Mo/Si, Cr/C, Sc/Cr, and W/B4C multilayers for transmission and reflection have also been designed and produced. The polarimeter mechanics is supported on a hexapod to simplify the alignment relative to photon beam. The instrument is designed so that it can be easily transferred between different beamlines.