Optimization of two-compartment-exchange-model analysis for dynamic contrast-enhanced mri incorporating bolus arrival time.
Nadav, Guy; Liberman, Gilad; Artzi, Moran; Kiryati, Nahum; Bashat, Dafna Ben.
Afiliação
  • Nadav G; Functional Brain Center, Wohl Institute for Advanced Imaging, Tel Aviv Sourasky Medical Center, Tel Aviv, Israel.
  • Liberman G; School of Electrical Engineering, Tel Aviv University, Tel Aviv, Israel.
  • Artzi M; Functional Brain Center, Wohl Institute for Advanced Imaging, Tel Aviv Sourasky Medical Center, Tel Aviv, Israel.
  • Kiryati N; Department of Chemical Physics, Weizmann Institute, Rehovot, Israel.
  • Bashat DB; Functional Brain Center, Wohl Institute for Advanced Imaging, Tel Aviv Sourasky Medical Center, Tel Aviv, Israel.
J Magn Reson Imaging ; 45(1): 237-249, 2017 01.
Article em En | MEDLINE | ID: mdl-27383624
ABSTRACT

PURPOSE:

To optimize the analysis of dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) under the two-compartment-exchange-model (2CXM) and to incorporate voxelwise bolus-arrival-time (BAT). MATERIALS AND

METHODS:

The accuracy of the pharmacokinetic (PK) parameters, extracted from 3T DCE-MRI using 2CXM, was tested under several conditions eight algorithms for data estimation; correction for BAT; using model selection; different temporal resolution and scan duration. Comparisons were performed on simulated data. The best algorithm was applied to seven patients with brain tumors or following stroke. The extracted perfusion parameters were compared to those of dynamic susceptibility contrast MRI (DSC-MRI).

RESULTS:

ACoPeD (AIF-corrected-perfusion-DCE-MRI), an analysis using a 2nd derivative regularized-spline and incorporating BAT, achieved the most accurate estimation in simulated data, mean-relative-error Fp , F, vp , ve 24.8%, 41.7%, 26.4%, 27.2% vs. 76.5%, 190.8%, 78.8%, 82.39% of the direct four parameters estimation (one-sided two-sample t-test, P < 0.001). Correction for BAT increased the estimation accuracy of the PK parameters by more than 30% and provided a supertemporal resolution estimation of the BAT (higher than the acquired resolution, mean-absolute-error 0.2 sec). High temporal resolution (∼2 sec) is required to avoid biased estimation of PK parameters, and long scan duration (∼20 min) is important for reliable permeability but not for perfusion estimations, mean-error-reduction E ∼12%, ve ∼6%. Using ACoPeD, PK values from normal-appearing white matter, gray matter, and lesion were extracted from patients. Preliminary results showed significant voxelwise correlations to DSC-MRI, between flow values in a patient following stroke (r = 0.49, P < 0.001), and blood volume in a patient with a brain tumor (r = 0.62, P < 0.001).

CONCLUSION:

This study proposes an optimized analysis method, ACoPeD, for tissue perfusion and permeability estimation using DCE-MRI, to be used in clinical settings. LEVEL OF EVIDENCE 1 J. Magn. Reson. Imaging 2017;45237-249.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Compostos Organometálicos / Neoplasias Encefálicas / Circulação Cerebrovascular / Angiografia por Ressonância Magnética / Meglumina / Modelos Cardiovasculares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Female / Humans / Male Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Israel

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Compostos Organometálicos / Neoplasias Encefálicas / Circulação Cerebrovascular / Angiografia por Ressonância Magnética / Meglumina / Modelos Cardiovasculares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Female / Humans / Male Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Israel