Your browser doesn't support javascript.
loading
Initial Technical Experience with the SMART Coil for the Embolization of Intracranial Aneurysms.
Spiotta, Alejandro M; Fargen, Kyle M; Lena, Jonathan; Chaudry, Imran; Turner, Raymond D; Turk, Aquilla S; Huddle, Daniel; Loy, David; Bellon, Richard; Frei, Donald.
Afiliação
  • Spiotta AM; Department of Neurosurgery, Medical University of South Carolina, Charleston, South Carolina, USA. Electronic address: spiotta@musc.edu.
  • Fargen KM; Department of Neurosurgery, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Lena J; Department of Neurosurgery, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Chaudry I; Department of Radiology, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Turner RD; Department of Neurosurgery, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Turk AS; Department of Radiology, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Huddle D; Department of Radiology Imaging Associates, Swedish Medical Center, Denver, Colorado, USA.
  • Loy D; Department of Radiology Imaging Associates, Swedish Medical Center, Denver, Colorado, USA.
  • Bellon R; Department of Radiology Imaging Associates, Swedish Medical Center, Denver, Colorado, USA.
  • Frei D; Department of Radiology Imaging Associates, Swedish Medical Center, Denver, Colorado, USA.
World Neurosurg ; 97: 80-85, 2017 Jan.
Article em En | MEDLINE | ID: mdl-27641255
BACKGROUND/OBJECTIVE: Traditional platinum coils used for cerebral aneurysm embolization have a uniform degree of softness throughout an individual coil's length. Recently, SMART Coils (Penumbra Inc., Alameda, California, USA) have been developed, which transition in softness along the length of their coils. We report the initial clinical results with this technology. METHODS: A retrospective study of all patients undergoing aneurysm coiling with SMART Coils at 2 centers was performed to evaluate the safety and efficacy of this new technology. Outcome and angiographic data were self-reported by individual treatment centers. RESULTS: Fifty-nine patients underwent treatment for cerebral aneurysms with SMART Coils (44% in the setting of subarachnoid hemorrhage). Mean aneurysm size was 5.9 ± 2 mm × 4.5 ± 2 mm. Mean neck size was 3.4 ± 1 mm. More than half (54.2%) of the patients underwent coiling exclusively with SMART Coils. The remainder used either framing or finishing coils of another type. About one third (33.9%) underwent balloon-assisted coiling, and 47.5% underwent stent-assisted coiling. There were no device malfunctions in any of the patients. Six (10.1%) minor complications without clinical sequelae occurred. Raymond I or II occlusion was achieved in 71.2%. CONCLUSIONS: Our results demonstrate adequate device safety in a variety of both ruptured and unruptured aneurysm locations. This design enables progressive changes in softness along the length of an individual coil, which offers several potential advantages in the clinical use. This initial investigation of SMART Coil technology demonstrates safety and efficacy in the treatment of a wide range of aneurysm sizes, locations, and morphologies in both ruptured and unruptured settings.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Stents / Aneurisma Intracraniano / Embolização Terapêutica Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male Idioma: En Revista: World Neurosurg Assunto da revista: NEUROCIRURGIA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Stents / Aneurisma Intracraniano / Embolização Terapêutica Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male Idioma: En Revista: World Neurosurg Assunto da revista: NEUROCIRURGIA Ano de publicação: 2017 Tipo de documento: Article