Performance of Vitek 2 for Antimicrobial Susceptibility Testing of Acinetobacter baumannii, Pseudomonas aeruginosa, and Stenotrophomonas maltophilia with Vitek 2 (2009 FDA) and CLSI M100S 26th Edition Breakpoints.
J Clin Microbiol
; 55(2): 450-456, 2017 02.
Article
em En
| MEDLINE
| ID: mdl-27881616
The performances of Vitek 2 AST-GN69 and AST-XN06 cards were compared to Clinical and Laboratory Standards Institute (CLSI) reference broth microdilution (BMD) for 99 isolates of Pseudomonas aeruginosa, 26 Acinetobacter baumannii isolates, and 11 Stenotrophomonas maltophilia isolates. In total, 15 antimicrobials were evaluated, with 11 for P. aeruginosa, 14 for A. baumannii, and 2 for S. maltophilia Categorical agreement (CA) was assessed using both Vitek 2 breakpoints and 2016 CLSI M100S 26th edition breakpoints. The essential agreement values for P. aeruginosa, A. baumannii, and S. maltophilia were 99.5%, 99.2%, and 100%, respectively. The CA values for P. aeruginosa, A. baumannii, and S. maltophilia were 94.1%, 92.7%, and 95.5%, respectively, by the Vitek 2 breakpoints, and 93.4%, 92.3%, and 95.5%, respectively, by the CLSI breakpoints. Overall, the Vitek 2 performance was comparable to that of BMD using both Vitek 2 breakpoints and 2016 CLSI M100S 26th edition breakpoints. Improved performance was noted for the reformulated piperacillin-tazobactam and imipenem found on the AST-GN69 card, with no very major or major errors noted when using the CLSI breakpoints.
Palavras-chave
Texto completo:
1
Bases de dados:
MEDLINE
Assunto principal:
Pseudomonas aeruginosa
/
Testes de Sensibilidade Microbiana
/
Stenotrophomonas maltophilia
/
Acinetobacter baumannii
/
Antibacterianos
Tipo de estudo:
Evaluation_studies
/
Guideline
Limite:
Humans
Idioma:
En
Revista:
J Clin Microbiol
Ano de publicação:
2017
Tipo de documento:
Article
País de afiliação:
Estados Unidos