Mechanisms of Undersensing by a Noise Detection Algorithm That Utilizes Far-Field Electrograms With Near-Field Bandpass Filtering.
Koneru, Jayanthi N; Swerdlow, Charles D; Ploux, Sylvain; Sharma, Parikshit S; Kaszala, Karoly; Tan, Alex Y; Huizar, Jose F; Vijayaraman, Pugazhendi; Kenigsberg, David; Ellenbogen, Kenneth A.
Afiliação
  • Koneru JN; Virginia Commonwealth University/Pauley Heart Center, Richmond, Virginia, USA.
  • Swerdlow CD; Cedars Sinai Heart Center, Los Angeles, California, USA.
  • Ploux S; Bordeaux University Hospital (CHU), Pessac, France.
  • Sharma PS; IHU Liryc, Electrophysiology and Heart Modeling Institute, Pessac-Bordeaux, France.
  • Kaszala K; Virginia Commonwealth University/Pauley Heart Center, Richmond, Virginia, USA.
  • Tan AY; Hunter Holmes McGuire Veterans Affairs Medical Center, Richmond, Virginia, USA.
  • Huizar JF; Hunter Holmes McGuire Veterans Affairs Medical Center, Richmond, Virginia, USA.
  • Vijayaraman P; Hunter Holmes McGuire Veterans Affairs Medical Center, Richmond, Virginia, USA.
  • Kenigsberg D; Geisinger Heart Institute, Wilkes-Barre, Pennsylvania, USA.
  • Ellenbogen KA; Westside Regional Medical Center, Plantation, Florida, USA.
J Cardiovasc Electrophysiol ; 28(2): 224-232, 2017 02.
Article em En | MEDLINE | ID: mdl-27957764
ABSTRACT

BACKGROUND:

Implantable cardioverter defibrillators (ICDs) must establish a balance between delivering appropriate shocks for ventricular tachyarrhythmias and withholding inappropriate shocks for lead-related oversensing ("noise"). To improve the specificity of ICD therapy, manufacturers have developed proprietary algorithms that detect lead noise. The SecureSenseTM RV Lead Noise discrimination (St. Jude Medical, St. Paul, MN, USA) algorithm is designed to differentiate oversensing due to lead failure from ventricular tachyarrhythmias and withhold therapies in the presence of sustained lead-related oversensing. METHODS AND

RESULTS:

We report 5 patients in whom appropriate ICD therapy was withheld due to the operation of the SecureSense algorithm and explain the mechanism for inhibition of therapy in each case. Limitations of algorithms designed to increase ICD therapy specificity, especially for the SecureSense algorithm, are analyzed.

CONCLUSION:

The SecureSense algorithm can withhold appropriate therapies for ventricular arrhythmias due to design and programming limitations. Electrophysiologists should have a thorough understanding of the SecureSense algorithm before routinely programming it and understand the implications for ventricular arrhythmia misclassification.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Algoritmos / Processamento de Sinais Assistido por Computador / Técnicas Eletrofisiológicas Cardíacas / Razão Sinal-Ruído Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Algoritmos / Processamento de Sinais Assistido por Computador / Técnicas Eletrofisiológicas Cardíacas / Razão Sinal-Ruído Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Aged / Female / Humans Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos