Your browser doesn't support javascript.
loading
Effect of Stem Size and Fixation Method on Mechanical Failure After Revision Total Knee Arthroplasty.
Fleischman, Andrew N; Azboy, Ibrahim; Fuery, Michael; Restrepo, Camilo; Shao, Hongyi; Parvizi, Javad.
Afiliação
  • Fleischman AN; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Azboy I; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Fuery M; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Restrepo C; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Shao H; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
  • Parvizi J; Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania.
J Arthroplasty ; 32(9S): S202-S208.e1, 2017 09.
Article em En | MEDLINE | ID: mdl-28559193
ABSTRACT

BACKGROUND:

Although the need for stemmed components is well accepted to improve mechanical survival in revision total knee arthroplasty (TKA), the ideal fixation method and stem design remain controversial.

METHODS:

We performed a retrospective review of 223 patients who underwent revision TKA in whom stemmed components had not been used previously and with a mean follow-up of 61.6 months, including 108 components with fully cemented stems and 316 components with "hybrid" press-fit stems.

RESULTS:

Based on a time to event model, risk for mechanical failure was equivalent for both cemented and hybrid stems (relative risk, 0.991; P = .98). Young age was the single greatest risk factor for mechanical failure (P = .006). Although there was a trend toward increased failure with cemented stems in patients aged <65 years, there was no significant difference in risk after accounting for covariates (relative risk, 1.4; P = .50). Intramedullary canal fill, not stem length or diameter, was the strongest predictor of failure with hybrid stems, and risk was reduced by 41.2% for each additional 10% canal fill.

CONCLUSION:

In conclusion, both cemented and hybrid modular stems are viable options in revision TKA. Surgeons should attempt to maximize canal filling of hybrid stems to obtain a solid press-fit. In addition, further studies are needed to evaluate the long-term survival of cemented stem fixation in young patients.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Reoperação / Artroplastia do Joelho / Prótese do Joelho Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Reoperação / Artroplastia do Joelho / Prótese do Joelho Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2017 Tipo de documento: Article