Fast and accurate multi-channel B1+ mapping based on the TIAMO technique for 7T UHF body MRI.
Brunheim, Sascha; Gratz, Marcel; Johst, Sören; Bitz, Andreas K; Fiedler, Thomas M; Ladd, Mark E; Quick, Harald H; Orzada, Stephan.
Afiliação
  • Brunheim S; Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany.
  • Gratz M; High Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany.
  • Johst S; Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany.
  • Bitz AK; High Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany.
  • Fiedler TM; Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany.
  • Ladd ME; Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Quick HH; Electromagnetic Theory and Applied Mathematics, Faculty of Electrical Engineering and Information Technology, FH Aachen-University of Applied Sciences, Aachen, Germany.
  • Orzada S; Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Magn Reson Med ; 79(5): 2652-2664, 2018 05.
Article em En | MEDLINE | ID: mdl-28994132
ABSTRACT

PURPOSE:

Current methods for mitigation of transmit field B1+ inhomogeneities at ultrahigh field (UHF) MRI by multi-channel radiofrequency (RF) shimming rely on accurate B1+ mapping. This can be time consuming when many RF channels have to be mapped for in vivo body MRI, where the B1 maps should ideally be acquired within a single breath-hold. Therefore, a new B1+ mapping technique (B1TIAMO) is proposed.

METHODS:

The performance of this technique is validated against an established method (DREAM) in phantom measurements for a cylindrical head phantom with an 8-channel transmit/receive (Tx/Rx) array. Furthermore, measurements for a 32-channel Tx/Rx remote array are conducted in a large body phantom and the |B1+| map reliability is validated against simulations of the transmit RF field distribution. Finally, in vivo results of this new mapping technique for human abdomen are presented.

RESULTS:

For the head phantom (8-channel Tx/Rx coil), the single |B1+| comparison between B1 TIAMO, the direct DREAM measurements, and simulation data showed good agreement with 10-19% difference. For the large body phantom (32-channel Tx/Rx coil), B1TIAMO matched the RF field simulations well.

CONCLUSION:

The results demonstrate the potential to acquire 32 accurate single-channel B1+ maps for large field-of-view body imaging within only a single breath-hold of 16 s at 7T UHF MRI. Magn Reson Med 792652-2664, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador Limite: Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador Limite: Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha