Your browser doesn't support javascript.
loading
Real-time beam monitoring for error detection in IMRT plans and impact on dose-volume histograms : A multi-center study.
Marrazzo, Livia; Arilli, Chiara; Pasler, Marlies; Kusters, Martijn; Canters, Richard; Fedeli, Luca; Calusi, Silvia; Casati, Marta; Talamonti, Cinzia; Simontacchi, Gabriele; Livi, Lorenzo; Pallotta, Stefania.
Afiliação
  • Marrazzo L; Medical Physic Unit, Careggi University Hospital, Largo Brambilla 3, 50134, Florence, Italy. livia.marrazzo@unifi.it.
  • Arilli C; Medical Physic Unit, Careggi University Hospital, Largo Brambilla 3, 50134, Florence, Italy.
  • Pasler M; Lake Constance Radiation Oncology Center, Singen-Friedrichshafen, Germany.
  • Kusters M; Department of Radiation Oncology, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Canters R; Department of Radiation Oncology, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Fedeli L; Department of Experimental and Clinical Biomedical Sciences "Mario Serio", University of Florence, Florence, Italy.
  • Calusi S; Department of Experimental and Clinical Biomedical Sciences "Mario Serio", University of Florence, Florence, Italy.
  • Casati M; Medical Physic Unit, Careggi University Hospital, Largo Brambilla 3, 50134, Florence, Italy.
  • Talamonti C; Medical Physic Unit, Careggi University Hospital, Largo Brambilla 3, 50134, Florence, Italy.
  • Simontacchi G; Department of Experimental and Clinical Biomedical Sciences "Mario Serio", University of Florence, Florence, Italy.
  • Livi L; Radiation Oncology Unit, Careggi University Hospital, Florence, Italy.
  • Pallotta S; Department of Experimental and Clinical Biomedical Sciences "Mario Serio", University of Florence, Florence, Italy.
Strahlenther Onkol ; 194(3): 243-254, 2018 03.
Article em En | MEDLINE | ID: mdl-29255923
PURPOSE: This study aimed to test the sensitivity of a transmission detector for online dose monitoring of intensity-modulated radiation therapy (IMRT) for detecting small delivery errors. Furthermore, the correlation of changes in detector output induced by small delivery errors with other metrics commonly employed to quantify the deviations between calculated and delivered dose distributions was investigated. METHODS: Transmission detector measurements were performed at three institutions. Seven types of errors were induced in nine clinical step-and-shoot (S&S) IMRT plans by modifying the number of monitor units (MU) and introducing small deviations in leaf positions. Signal reproducibility was investigated for short- and long-term stability. Calculated dose distributions were compared in terms of γ passing rates and dose-volume histogram (DVH) metrics (e.g., Dmean, Dx%, Vx%). The correlation between detector signal variations, γ passing rates, and DVH parameters was investigated. RESULTS: Both short- and long-term reproducibility was within 1%. Dose variations down to 1 MU (∆signal 1.1 ± 0.4%) as well as changes in field size and positions down to 1 mm (∆signal 2.6 ± 1.0%) were detected, thus indicating high error-detection sensitivity. A moderate correlation of detector signal was observed with γ passing rates (R2 = 0.57-0.70), while a good correlation was observed with DVH metrics (R2 = 0.75-0.98). CONCLUSION: The detector is capable of detecting small delivery errors in MU and leaf positions, and is thus a highly sensitive dose monitoring device for S&S IMRT for clinical practice. The results of this study indicate a good correlation of detector signal with DVH metrics; therefore, clinical action levels can be defined based on the presented data.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Sistemas Computacionais / Planejamento da Radioterapia Assistida por Computador / Monitoramento de Radiação / Radioterapia de Intensidade Modulada / Erros de Configuração em Radioterapia Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies Limite: Humans Idioma: En Revista: Strahlenther Onkol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Sistemas Computacionais / Planejamento da Radioterapia Assistida por Computador / Monitoramento de Radiação / Radioterapia de Intensidade Modulada / Erros de Configuração em Radioterapia Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies Limite: Humans Idioma: En Revista: Strahlenther Onkol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Itália