Your browser doesn't support javascript.
loading
Management of a leadless pacemaker recall via extraction in a patient with a history of multiple previous device implants.
Curnis, Antonio; Cerini, Manuel; Inama, Lorenza; Vassanelli, Francesca; Mariggiò, Davide; Salghetti, Francesca; Mitacchione, Gianfranco; Giacopelli, Daniele; Bontempi, Luca.
Afiliação
  • Curnis A; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.
  • Cerini M; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.
  • Inama L; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.
  • Vassanelli F; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.
  • Mariggiò D; Medtronic Italia, Milano, Italy.
  • Salghetti F; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.
  • Mitacchione G; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy; Department of Heart and Vessels, Ospedale di Circolo e Fondazione Macchi, University of Insubria, Varese, Italy.
  • Giacopelli D; Biotronik Italia, Vimodrone, Italy.
  • Bontempi L; Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy. Electronic address: bontempiluca@libero.it.
J Electrocardiol ; 53: 5-7, 2019.
Article em En | MEDLINE | ID: mdl-30572227
We describe a management strategy of a leadless pacemaker recall (LP) via extraction in a pacemaker-dependent patient with a history of multiple previous device implants. We performed a two-step procedure. First, we implanted a second LP in the right ventricle. Then, 3 days later for concerns about the stability of the newly-implanted device, we retrieved the first LP, 3 years after its original implant. The patient was hemodynamically stable throughout the two procedures and no adverse event occurred in the 3 days of coexistence of the two LPs. Although the limitation of a single case experience, this approach may be taken into consideration when facing similar high-risk cases.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Remoção de Dispositivo / Endocardite Bacteriana / Bloqueio Atrioventricular Limite: Aged80 / Humans / Male Idioma: En Revista: J Electrocardiol Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Remoção de Dispositivo / Endocardite Bacteriana / Bloqueio Atrioventricular Limite: Aged80 / Humans / Male Idioma: En Revista: J Electrocardiol Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Itália