Your browser doesn't support javascript.
loading
Simultaneous proton magnetic resonance fingerprinting and sodium MRI.
Yu, Zidan; Madelin, Guillaume; Sodickson, Daniel K; Cloos, Martijn A.
Afiliação
  • Yu Z; Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York.
  • Madelin G; Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University School of Medicine, New York, New York.
  • Sodickson DK; The Sackler Institute of Graduate Biomedical Sciences, New York University School of Medicine, New York, New York.
  • Cloos MA; Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York.
Magn Reson Med ; 83(6): 2232-2242, 2020 06.
Article em En | MEDLINE | ID: mdl-31746048
ABSTRACT

PURPOSE:

The goal of this work is to demonstrate a method for the simultaneous acquisition of proton multiparametric maps (T1 , T2 , and proton density) and sodium density images in 1 single scan. We hope that the development of such capabilities will help to ease the implementation of sodium MRI in clinical trials and provide more opportunities for researchers to investigate metabolism through sodium MRI.

METHODS:

We developed a sequence based on magnetic resonance fingerprinting (MRF), which contains interleaved proton (1 H) and sodium (23 Na) excitations followed by a simultaneous center-out radial readout for both nuclei. The receive chain of a 7T scanner was modified to enable simultaneous acquisition of 1 H and 23 Na signal. The obtained signal-to-noise ratio (SNR) was evaluated, and the accuracy of both proton T1 , T2 , and B1+ and sodium density maps were verified in phantoms. Finally, the method was demonstrated in 2 healthy subjects.

RESULTS:

The SNR obtained using the simultaneous measurement was almost identical to single-nucleus measurements (<1% change). Similarly, the proton T1 and T2 maps remained stable (normalized root mean square error in T1 ≈ 2.2%, in T2 ≈ 1.4%, and B1+ ≈ 5.4%), which indicates that the proposed sequence and hardware have no significant effects on the signal from either nucleus. In vivo measurements corroborated these results and demonstrated the feasibility of our method for in vivo application.

CONCLUSIONS:

With the proposed approach, we were able to simultaneously acquire sodium density images in addition to proton T1 , T2 , and B1+ maps as well as proton density images.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Prótons / Sódio Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Prótons / Sódio Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article