The rotational fixator: A new device.
Omar Pacha, T; Khalifa, A; Graulich, T; Alaidarous, H; Omar, M; Krettek, C; Stubig, T.
Afiliação
  • Omar Pacha T; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
  • Khalifa A; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
  • Graulich T; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
  • Alaidarous H; Trauma Department, Hessing Klinik, Hessingstr. 17, Bavaria, Augsburg, 86199, Germany.
  • Omar M; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
  • Krettek C; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
  • Stubig T; Trauma Department, Hannover Medical School(MHH), MHH, Carl-Neuberg-Str.1, Lower Saxony, 30625, Hannover, Germany.
J Orthop ; 19: 150-152, 2020.
Article em En | MEDLINE | ID: mdl-32025123
ABSTRACT

OBJECTIVES:

Malrotation after surgical treatment of femoral shaft fractures is a common problem and often leads to follow-up procedures with uncertain outcome. The aim of this study is the validation of a new device (Rotational Fixator) to perform the correction safely and accurately.

METHODS:

In an in-vitro study, we tested the Rotational Fixator on 21 corpse bones against a commercially available standard goniometer for measurement inaccuracies. For this purpose, we varied the rotation width from 10 to 30° in inside and outside rotation.

RESULTS:

We found a small measurement inaccuracy of 1-2° with increasing rotation. The smallest differences are found at 10° IR with 0.9524° (SD ± 1.0713; p = 0.001) difference and 10° ER with at 0.5952° (SD ± 0.6823; p = 0.001) difference and increase up to 30° (IR 1.6667°, SD ± 1.7121, p < 0.000/ER 1.5000°, SD ± 1.0488, p < 0.000).

CONCLUSIONS:

The measurement results of the device show a constant deviation from the gold standard but are constant in the measurement error and slightly in relation to the desired correction range, so that a further review of the device and further testing in in vivo studies makes sense. LEVELS OF EVIDENCE Level 3.