Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.
Ahmadi, Ali; Klein, Ran; Lewin, Howard C; Beanlands, Rob S B; deKemp, Robert A.
Afiliação
  • Ahmadi A; Division of Cardiology, University of Ottawa Heart Institute, National Cardiac PET Centre, 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.
  • Klein R; Division of Nuclear Medicine, The Ottawa Hospital, Ottawa, ON, Canada.
  • Lewin HC; Cardiac Imaging Nuclear Associates, Los Angeles, CA, USA.
  • Beanlands RSB; Division of Cardiology, University of Ottawa Heart Institute, National Cardiac PET Centre, 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.
  • deKemp RA; Division of Cardiology, University of Ottawa Heart Institute, National Cardiac PET Centre, 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada. RAdeKemp@ottawaheart.ca.
J Nucl Cardiol ; 27(5): 1728-1738, 2020 10.
Article em En | MEDLINE | ID: mdl-32436115
ABSTRACT

INTRODUCTION:

Strontium-82/Rubidium-82 (82Sr/82Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the 82Rb isotope yield and production efficiency of two FDA-approved 82Sr/82Rb generators were compared.

METHODS:

N = 515 sequential daily quality assurance (QA) reports from 9 CardioGen-82® and 9 RUBY-FILL® generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL® system to determine an empirical correction-factor used to convert CardioGen-82® daily QA values of 82Rb activity (dose-calibrator 'maximum' of 50 mL elution at 50 mL·min-1) to RUBY-FILL® equivalent values (integrated 'total' of 35 mL elution at 20 mL·min-1). The generator yield (82Rb) and production efficiency (82Rb yield/82Sr parent activity) were measured and compared after this conversion to a common scale.

RESULTS:

At the start of clinical use, the system reported 82Rb activity from daily QA was lower for CardioGen-82® vs RUBY-FILL® (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, P < 0.001) despite having similar 82Sr activity. Dose-calibrator 'maximum' (CardioGen-82®) values were found to under-estimate the integrated 'total' (RUBY-FILL®) activity by ~ 24% at 50 mL·min-1. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82® efficiency remained slightly lower than the RUBY-FILL® system on average (88 ± 4% vs 95 ± 4%, P < 0.001). The efficiency of 82Rb production improved for both systems over the respective periods of clinical use.

CONCLUSIONS:

82Rb generator yield was significantly under-estimated using the CardioGen-82® vs RUBY-FILL® daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated 82Rb production efficiency of the CardioGen-82® system was ~ 7% lower than RUBY-FILL® over the full period of clinical use.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Geradores de Radionuclídeos / Radioisótopos de Estrôncio / Radioisótopos de Rubídio / Tomografia por Emissão de Pósitrons / Imagem de Perfusão do Miocárdio / Cardiopatias Limite: Humans Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Geradores de Radionuclídeos / Radioisótopos de Estrôncio / Radioisótopos de Rubídio / Tomografia por Emissão de Pósitrons / Imagem de Perfusão do Miocárdio / Cardiopatias Limite: Humans Idioma: En Revista: J Nucl Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Canadá