System performance evaluation of the cobas t 711 and cobas t 511 coagulation analyzers in routine laboratory settings.
Kitchen, Steve; de Maat, Moniek; Nagler, Michael; Jones, Robert; Lowe, Anna; Burden, John; Groneberg, Kai; Rozsnyai, Gergely.
Afiliação
  • Kitchen S; Sheffield Haemostasis and Thrombosis Centre, Sheffield, UK.
  • de Maat M; Department of Haematology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands.
  • Nagler M; University Institute of Clinical Chemistry, Inselspital, Bern University Hospital, Bern.
  • Jones R; Sheffield Haemostasis and Thrombosis Centre, Sheffield, UK.
  • Lowe A; Sheffield Haemostasis and Thrombosis Centre, Sheffield, UK.
  • Burden J; Roche Diagnostics International Ltd, Rotkreuz, Switzerland.
  • Groneberg K; Roche Diagnostics International Ltd, Rotkreuz, Switzerland.
  • Rozsnyai G; Roche Diagnostics International Ltd, Rotkreuz, Switzerland.
Blood Coagul Fibrinolysis ; 31(7): 459-468, 2020 Oct.
Article em En | MEDLINE | ID: mdl-32826393
ABSTRACT
Utility of coagulation analyzers in real-world settings depends on characteristics that are often not studied comprehensively. This study aimed to investigate the analytical performance, system functionality, practicability, consistency and throughput of two new automated coagulation analyzers in routine laboratory practice. Real-world settings were simulated in three major European hemostasis laboratories and multiple assays were performed in anonymized plasma samples in parallel with routine clinical practice on the cobas t 711 (high-throughput) and cobas t 511 (mid-throughput) analyzers using activated partial thromboplastin time (aPTT), aPTT Lupus, aPTT Screen, Antithrombin (AT), D-Dimer, Fibrinogen, Prothrombin Time (PT)-derived Fibrinogen, PT Owren, PT Rec (recombinant human thromboplastin reagent) and Thrombin Time assays. Precision was tested in a 21-day experiment and accuracy was compared with reference methods of the same laboratory. A number of experiments simulated challenging real-life situations. Pearson's correlation coefficient was more than 0.98 in all assays. Across assays, coefficients of variation ranged from 0.0 to 1.5% for intermediate precision; 0.2 to 3.0% for repeatability and 0.4 to 3.7% for total precision. Good between-run comparability was seen when testing samples under random conditions. Calculated maximum throughput was 197 and 387-402 tests/h for the cobas t 511 and 711 analyzers, respectively. Practicability met or exceeded user expectations in 98% of cases. In a simulated real-life setting of three major laboratories, the new cobas t 511 and cobas t 711 coagulation analyzers demonstrated a good functionality, practicability and performance and the throughput was high.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Tempo de Tromboplastina Parcial / Testes de Coagulação Sanguínea / Laboratórios Limite: Humans Idioma: En Revista: Blood Coagul Fibrinolysis Assunto da revista: ANGIOLOGIA / HEMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Tempo de Tromboplastina Parcial / Testes de Coagulação Sanguínea / Laboratórios Limite: Humans Idioma: En Revista: Blood Coagul Fibrinolysis Assunto da revista: ANGIOLOGIA / HEMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Reino Unido