Standardized fluoroscopy-guided implantation technique enables optimal electrode placement in sacral neuromodulation: a cadaver study.
Müller, C; Reissig, L F; Argeny, S; Weninger, W J; Riss, S.
Afiliação
  • Müller C; Division of General Surgery, Department of Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090, Vienna, Austria.
  • Reissig LF; Division of Anatomy, MIC Medical University of Vienna, Vienna, Austria.
  • Argeny S; Division of General Surgery, Department of Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090, Vienna, Austria.
  • Weninger WJ; Division of Anatomy, MIC Medical University of Vienna, Vienna, Austria.
  • Riss S; Division of General Surgery, Department of Surgery, Medical University of Vienna, Währinger Gürtel 18-20, 1090, Vienna, Austria. stefan.riss@meduniwien.ac.at.
Tech Coloproctol ; 25(2): 215-221, 2021 02.
Article em En | MEDLINE | ID: mdl-33211205
ABSTRACT

BACKGROUND:

Sacral neuromodulation (SNM) is an established treatment option for patients with faecal incontinence. The location of the stimulating electrode is considered to be essential for treatment success. The purpose of this study was to evaluate the position of SNM electrodes after using a standardized fluoroscopy-guided implantation technique.

METHODS:

For this cadaver study, SNM electrodes were implanted bilaterally in 5 lower body specimens. The lower edge of the sacroiliac joint and the medial edge of the sacral foramina were marked using fluoroscopy to draw an 'H' with the crossing points identifying S3. After electrode placement the pelvis was dissected to describe the exact position of the SNM electrodes.

RESULTS:

The electrodes were inserted at an angle with a median degree measure of 60° (range 50-65°) to the skin, with a median distance of 9 mm (range 0-13 mm) from the S3 marking. All electrodes entered the third sacral foramen. The median distance of the electrodes to the sacral nerve was 0 mm (range 0-3 mm) for the most proximal, 0.5 mm (range 0-5 mm) for the second, 2.25 mm (range 0-11 mm) for the third and 1.75 mm (range 0-16 mm) for the most distant electrode. There was neither a significant difference in the proximity of the electrodes to the nerve between the right and left side (proximal to distal electrode p = 0.18, p = 0.16, p = 0.07, p = 0.07) nor between male and female cadavers (p = 0.25, p = 0.21, p = 0.66, p = 0.66).

CONCLUSIONS:

A standardized fluoroscopy-guided implantation technique enables a close contact between electrode and nerve. This can potentially result in an improved clinical outcome.