Comparison of Inline R2* MRI versus FerriScan for liver iron quantification in patients on chelation therapy for iron overload: preliminary results.
Eur Radiol
; 31(12): 9296-9305, 2021 Dec.
Article
em En
| MEDLINE
| ID: mdl-34041571
ABSTRACT
OBJECTIVES:
MRI quantification of liver iron concentration (LIC) using R2 or R2* relaxometry requires offline post-processing causing reporting delays, administrative overhead, and added costs. A prototype 3D multi-gradient-echo pulse sequence, with inline post-processing, allows immediate calculation of LIC from an R2* map (inline R2*-LIC) without offline processing. We compared inline R2*-LIC to FerriScan and offline R2* calibration methods.METHODS:
Forty patients (25 women, 15 men; age 18-82 years), prospectively underwent FerriScan and the prototype sequence, which produces two R2* maps, with and without fat modeling, as well as an inline R2*-LIC map derived from the R2* map with fat modeling, with informed consent. For each map, the following contours were drawn ROIs, whole-axial-liver contour, and an exact copy of contour utilized by FerriScan. LIC values from the FerriScan report and those calculated using an alternative R2 calibration were the reference standards. Results were compared using Pearson and interclass correlation coefficients (PCC, ICC), linear regression, Bland-Altman analysis, and estimation of area under the receiver operator curve (ROC-AUC).RESULTS:
Inline R2*-LIC demonstrated good agreement with the reference standards. Compared to FerriScan, inline R2*-LIC with whole-axial-liver contour, ROIs, and FerriScan contour demonstrated PCC of 94.8%, 94.8%, and 92%; ICC 93%, 92.7%, and 90.2%; regression slopes 1.004, 0.974, and 1.031; mean bias 5.54%, 10.91%, and 0.36%; and ROC-AUC estimates 0.903, 0.906, and 0.890 respectively. Agreement was maintained when adjusted for sex, age, diagnosis, liver fat content, and fat-water swap.CONCLUSION:
Inline R2*-LIC provides robust and comparable quantification of LIC compared to FerriScan, without the need for offline post-processing. KEY POINTS ⢠In patients being treated for iron overload with chelation therapy, liver iron concentration (LIC) is regularly assessed in order to monitor and adjust therapy. ⢠Magnetic resonance imaging (MRI) is commonly used to quantify LIC. Several R2 and R2* methods are available, all of which require offline post-processing. ⢠A novel R2* MRI method allows for immediate calculation of LIC and provides comparable quantification of LIC to the FerriScan and recently published alternative R2* methods.Palavras-chave
Texto completo:
1
Bases de dados:
MEDLINE
Assunto principal:
Sobrecarga de Ferro
/
Ferro
Tipo de estudo:
Guideline
Limite:
Adolescent
/
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Eur Radiol
Assunto da revista:
RADIOLOGIA
Ano de publicação:
2021
Tipo de documento:
Article
País de afiliação:
Canadá