A standarized protocol for evaluation of large extracellular vesicles using the attune™ NXT system.
J Immunol Methods
; 499: 113170, 2021 12.
Article
em En
| MEDLINE
| ID: mdl-34662560
ABSTRACT
Extracellular vesicles (EVs) represent a promising biomarker in several medical areas. Flow cytometry (FC) is one of the most widely-used methods to characterize EVs, providing quantitative information and determination of EV subtypes. EV evaluation represents a challenge as no standardized methods are available to facilitate assessment across different research centers. This is principally because their size falls below the detection limit of most standard flow cytometers and a thorough optimization process is required to ensure instrument-specific sensitivity. We provide an overview of a standardized method to evaluate large EVs using the Attune™ Nxt Acoustic Focusing Flow Cytometer system (Thermo Fisher Scientific).
Palavras-chave
Texto completo:
1
Bases de dados:
MEDLINE
Assunto principal:
Citometria de Fluxo
/
Vesículas Extracelulares
Limite:
Humans
Idioma:
En
Revista:
J Immunol Methods
Ano de publicação:
2021
Tipo de documento:
Article
País de afiliação:
Austrália