Your browser doesn't support javascript.
loading
FT-ICR Mass Spectrometry Imaging at Extreme Mass Resolving Power Using a Dynamically Harmonized ICR Cell with 1ω or 2ω Detection.
Tiquet, Mathieu; La Rocca, Raphaël; Kirnbauer, Stefan; Zoratto, Samuele; Van Kruining, Daan; Quinton, Loïc; Eppe, Gauthier; Martinez-Martinez, Pilar; Marchetti-Deschmann, Martina; De Pauw, Edwin; Far, Johann.
Afiliação
  • Tiquet M; Mass Spectrometry Laboratory, MolSys Research Unit, University of Liège, Allée de la Chimie 6-Quartier Agora, 4000 Liège, Belgium.
  • La Rocca R; Mass Spectrometry Laboratory, MolSys Research Unit, University of Liège, Allée de la Chimie 6-Quartier Agora, 4000 Liège, Belgium.
  • Kirnbauer S; Institute of Chemical Technologies and Analytics, TU Wien (Vienna University of Technology), Getreidemarkt 9/164, 1060 Vienna, Austria.
  • Zoratto S; Austrian Cluster for Tissue Regeneration, TU Wien (Vienna University of Technology), Getreidemarkt 9/164, 1060 Vienna, Austria.
  • Van Kruining D; Institute of Chemical Technologies and Analytics, TU Wien (Vienna University of Technology), Getreidemarkt 9/164, 1060 Vienna, Austria.
  • Quinton L; Austrian Cluster for Tissue Regeneration, TU Wien (Vienna University of Technology), Getreidemarkt 9/164, 1060 Vienna, Austria.
  • Eppe G; Christian Doppler Laboratory for Skin Multimodal Imaging of Aging and Senescence, TU Wien (Vienna University of Technology), Getreidemarkt 9/164, 1060 Vienna, Austria.
  • Martinez-Martinez P; Department of Psychiatry and Neuropsychology, School for Mental Health and Neuroscience, Maastricht University, Universiteitssingel 50, 6229ER Maastricht, the Netherlands.
  • Marchetti-Deschmann M; Mass Spectrometry Laboratory, MolSys Research Unit, University of Liège, Allée de la Chimie 6-Quartier Agora, 4000 Liège, Belgium.
  • De Pauw E; Mass Spectrometry Laboratory, MolSys Research Unit, University of Liège, Allée de la Chimie 6-Quartier Agora, 4000 Liège, Belgium.
  • Far J; Department of Psychiatry and Neuropsychology, School for Mental Health and Neuroscience, Maastricht University, Universiteitssingel 50, 6229ER Maastricht, the Netherlands.
Anal Chem ; 94(26): 9316-9326, 2022 07 05.
Article em En | MEDLINE | ID: mdl-35604839
ABSTRACT
MALDI mass spectrometry imaging (MALDI MSI) is a powerful analytical method for achieving 2D localization of compounds from thin sections of typically but not exclusively biological samples. The dynamically harmonized ICR cell (ParaCell) was recently introduced to achieve extreme spectral resolution capable of providing the isotopic fine structure of ions detected in complex samples. The latest improvement in the ICR technology also includes 2ω detection, which significantly reduces the transient time while preserving the nominal mass resolving power of the ICR cell. High-resolution MS images acquired on FT-ICR instruments equipped with 7T and 9.4T superconducting magnets and the dynamically harmonized ICR cell operating at suboptimal parameters suffered severely from the pixel-to-pixel shifting of m/z peaks due to space-charge effects. The resulting profile average mass spectra have depreciated mass measurement accuracy and mass resolving power under the instrument specifications that affect the confidence level of the identified ions. Here, we propose an analytical workflow based on the monitoring of the total ion current to restrain the pixel-to-pixel m/z shift. Adjustment of the laser parameters is proposed to maintain high spectral resolution and mass accuracy measurement within the instrument specifications during MSI analyses. The optimized method has been successfully employed in replicates to perform high-quality MALDI MS images at resolving power (FWHM) above 1,000,000 in the lipid mass range across the whole image for superconducting magnets of 7T and 9.4T using 1 and 2ω detection. Our data also compare favorably with MALDI MSI experiments performed on higher-magnetic-field superconducting magnets, including the 21T MALDI FT-ICR prototype instrument of the NHMFL group at Tallahassee, Florida.
Assuntos

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Diagnóstico por Imagem / Ciclotrons Tipo de estudo: Diagnostic_studies Idioma: En Revista: Anal Chem Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Bélgica

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Diagnóstico por Imagem / Ciclotrons Tipo de estudo: Diagnostic_studies Idioma: En Revista: Anal Chem Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Bélgica