Fast bound and pore water mapping of cortical bone with arbitrary slice oriented two-dimensional ultra-short echo time.
Harkins, Kevin D; Ketsiri, Thammathida; Nyman, Jeffry S; Does, Mark D.
Afiliação
  • Harkins KD; Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Ketsiri T; Institute of Image Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Nyman JS; Biomedical Engineering, Vanderbilt University, Nashville, Tennessee, USA.
  • Does MD; Institute of Image Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Magn Reson Med ; 89(2): 767-773, 2023 02.
Article em En | MEDLINE | ID: mdl-36226656
ABSTRACT

PURPOSE:

Extend fast, two-dimensional (2D) methods of bound and pore water mapping in bone to arbitrary slice orientation.

METHODS:

To correct for slice profile artifacts caused by gradient errors of half pulse 2D ultra-short echo time (UTE), we developed a library of predistorted gradient waveforms that can be used to interpolate optimized gradient waveforms for 2D UTE slice selection. We also developed a method to estimate and correct for a bulk phase difference between the two half pulse excitations used for 2D UTE signal excitation. Bound water images were acquired in three healthy subjects with adiabatic inversion recovery prepared 2D UTE, while pore water images were acquired after short-T2 signals were suppressed with double adiabatic inversion recovery preparation. The repeatability of bound and pore water imaging with 2D UTE was tested by repeating acquisitions after repositioning.

RESULTS:

The library-based interpolation of optimized slice select gradient waveforms combined with the method to estimate bulk phase between two excitations provided compact slice profiles for half pulse excited 2D UTE. Quantitative bound and pore water values were highly repeatable-the pooled SD of bound water across all three subjects was 0.38 mol 1 $$ {}^1 $$ H/L, while pooled SD of pore water was 0.30 mol 1 $$ {}^1 $$ H/L.

CONCLUSION:

Fast, quantitative, 2D UTE-based bound and pore water images can be acquired at arbitrary oblique orientations after correcting for errors in the slice select gradient waveform and bulk phase shift between the two half acquisitions.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Água Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos