Error on the stopping power ratio of ERKODENT's mouthpiece for head and neck carbon ion radiotherapy treatment.
Lee, Sung Hyun; Kanai, Takayuki; Souda, Hikaru; Miyasaka, Yuya; Chai, Hongbo; Ono, Takuya; Yamazawa, Yoshifumi; Suzuki, Koji; Sato, Azusa; Katsumata, Masashi; Iwai, Takeo.
Afiliação
  • Lee SH; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Kanai T; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Souda H; Department of Radiation Oncology, Tokyo Women's Medical University, Shinjuku, Tokyo, Japan.
  • Miyasaka Y; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Chai H; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Ono T; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Yamazawa Y; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Suzuki K; Department of Heavy Particle Medical Science, Graduate School of Medical Science, Yamagata University, Iidanishi, Yamagata, Japan.
  • Sato A; Department of Radiology, Yamagata University Hospital, Iidanishi, Yamagata, Japan.
  • Katsumata M; Department of Radiology, Yamagata University Hospital, Iidanishi, Yamagata, Japan.
  • Iwai T; Accelerator Engineering Corporation, Inage, Chiba, Japan.
J Appl Clin Med Phys ; 24(5): e13987, 2023 May.
Article em En | MEDLINE | ID: mdl-37018016
ABSTRACT
The errors on the stopping power ratio (SPR) of mouthpiece samples from ERKODENT were evaluated. Erkoflex and Erkoloc-pro from ERKODENT and samples that combined Erkoflex and Erkoloc-pro were computed tomography (CT)-scanned using head and neck (HN) protocol at the East Japan Heavy Ion Center (EJHIC), and the values were averaged to obtain the CT number. The integral depth dose of the Bragg curve with and without these samples was measured for 292.1, 180.9, and 118.8 MeV/u of the carbon-ion pencil beam using an ionization chamber with concentric electrodes at the horizontal port of the EJHIC. The average value of the water equivalent length (WEL) of each sample was obtained from the difference between the range of the Bragg curve and the thickness of the sample. To calculate the difference between the theoretical and measured values, the theoretical CT number and SPR value of the sample were calculated using the stoichiometric calibration method. Compared with the Hounsfield unit (HU)-SPR calibration curve used at the EJHIC, the SPR error on each measured and theoretical value was calculated. The WEL value of the mouthpiece sample had an error of approximately 3.5% in the HU-SPR calibration curve. From this error, it was evaluated that for a mouthpiece with a thickness of 10 mm, a beam range error of approximately 0.4 mm can occur, and for a mouthpiece with a thickness of 30 mm, a beam range error of approximately 1 mm can occur. For a beam passing through the mouthpiece in HN treatment, it would be practical to consider a mouthpiece margin of 1 mm to avoid beam range errors if ions pass through the mouthpiece.