Ultrafast inactivation of SARS-CoV-2 by 254-nm UV-C irradiation on porous and non-porous media of medical interest using an omnidirectional chamber.
Maubert, Bertrand; Theillière, Camille; Jany, Prescillia; Bourlet, Thomas; Deschamps, Jérôme; Pozzetto, Bruno; Singh, Fateh; Gadea, Emilie.
Afiliação
  • Maubert B; Laboratoire de Biologie, Centre Hospitalier Emile Roux, 43000, Le Puy en Velay, France.
  • Theillière C; Unité de Recherche Clinique, Centre Hospitalier Emile Roux, 43000, Le Puy en Velay, France.
  • Jany P; CIRI, Centre International de Recherche en Infectiologie, GIMAP Team, Univ St-Etienne, INSERM U1111, CNRS UMR5308, ENS de Lyon, UCBL1, Univ Lyon, 42023, Saint-Etienne, France.
  • Bourlet T; Unité de Recherche Clinique, Centre Hospitalier Emile Roux, 43000, Le Puy en Velay, France.
  • Deschamps J; CIRI, Centre International de Recherche en Infectiologie, GIMAP Team, Univ St-Etienne, INSERM U1111, CNRS UMR5308, ENS de Lyon, UCBL1, Univ Lyon, 42023, Saint-Etienne, France.
  • Pozzetto B; Service des Agents Infectieux, Centre Hospitalier Universitaire de Saint-Étienne, 42055, Saint-Etienne, France.
  • Singh F; ON-LIGHT SAS, SMO Biopole Clermont-Limagne, 63360, Saint Beauzire, France.
  • Gadea E; CIRI, Centre International de Recherche en Infectiologie, GIMAP Team, Univ St-Etienne, INSERM U1111, CNRS UMR5308, ENS de Lyon, UCBL1, Univ Lyon, 42023, Saint-Etienne, France.
Sci Rep ; 13(1): 12648, 2023 08 04.
Article em En | MEDLINE | ID: mdl-37542073
ABSTRACT
Covid-19 has spurred a renewed interest in decontamination techniques for air, objects and surfaces. Beginning in 2020, urgent effort was done to permit the reuse of UV-C for inactivating SARS-CoV-2. However, those studies diverged widely on the dose necessary to reach this goal; until today, the real value of the sensitivity of the virus to a 254-nm illumination is not known precisely. In this study, decontamination was performed in an original UV-C large decontamination chamber (UVCab, ON-LIGHT, France) delivering an omnidirectional irradiation with an average dose of 50 mJ/cm2 in 60 s. Viral inactivation was checked by both cell culture and PCR test. SARS-CoV-2 was inactivated by UV-C light within 3 s on both porous (disposable gown) and non-porous (stainless steel and apron) surfaces. For the porous surface, an irradiation of 5 min was needed to achieve a completely negative PCR signal. The Z value estimating the sensitivity of SARS-CoV-2 to UV-C in the experimental conditions of our cabinet was shown to be > 0.5820 m2/J. These results illustrate the ability of this apparatus to inactivate rapidly and definitively high loads of SARS-CoV-2 deposited on porous or non-porous supports and opens new perspectives on material decontamination using UV-C.