Your browser doesn't support javascript.
loading
An anthropomorphic thyroid phantom for ultrasound-guided radiofrequency ablation of nodules.
Boers, Tim; Brink, Wyger; Bianchi, Leonardo; Saccomandi, Paola; van Hespen, Johan; Wennemars, Germen; Braak, Sicco; Versluis, Michel; Manohar, Srirang.
Afiliação
  • Boers T; Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Brink W; Magnetic Detection and Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Bianchi L; Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Saccomandi P; Department of Mechanical Engineering, Politecnico di Milano, Milan, Italy.
  • van Hespen J; Department of Mechanical Engineering, Politecnico di Milano, Milan, Italy.
  • Wennemars G; Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Braak S; Magnetic Detection and Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Versluis M; Department of Radiology, Ziekenhuisgroep Twente, Almelo, the Netherlands.
  • Manohar S; Physics of Fluids group, TechMed Centre, University of Twente, Enschede, the Netherlands.
Med Phys ; 51(2): 826-838, 2024 Feb.
Article em En | MEDLINE | ID: mdl-38141047
ABSTRACT

BACKGROUND:

Needle-based procedures, such as fine needle aspiration and thermal ablation, are often applied for thyroid nodule diagnosis and therapeutic purposes, respectively. With blood vessels and nerves nearby, these procedures can pose risks in damaging surrounding critical structures.

PURPOSE:

The development and validation of innovative strategies to manage these risks require a test object with well-characterized physical properties. For this work, we focus on the application of ultrasound-guided thermal radiofrequency ablation.

METHODS:

We have developed a single-use anthropomorphic phantom mimicking the thyroid and surrounding anatomical and physiological structures that are relevant to ultrasound-guided thermal ablation. The phantom was composed of a mixture of polyacrylamide, water, and egg white extract and was cast using molds in multiple steps. The thermal, acoustical, and electrical characteristics were experimentally validated. The ablation zones were analyzed via non-destructive T2 -weighted magnetic resonance imaging scans utilizing the relaxometry changes of coagulated egg albumen, and the temperature distribution was monitored using an array of fiber Bragg grating sensors.

RESULTS:

The physical properties of the phantom were verified both on ultrasound as well as in terms of the phantom response to thermal ablation. The final temperature achieved (92°C), the median percentage of the nodule ablated (82.1%), the median volume ablated outside the nodule (0.8 mL), and the median number of critical structures affected (0) were quantified.

CONCLUSION:

An anthropomorphic phantom that can provide a realistic model for development and training in ultrasound-guided needle-based thermal interventions for thyroid nodules has been presented.