Safety and feasibility of a novel total percutaneous post-closure technique after bedside Impella decannulation in patients with cardiogenic shock.
Suzuki, Yuichi; Mogi, Satoshi; Mizuno, Yusuke; Sato, Ryota; Satoh, Terumori; Ohtani, Hayato; Saotome, Masao; Maekawa, Yuichiro.
Afiliação
  • Suzuki Y; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Mogi S; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Mizuno Y; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Sato R; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Satoh T; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Ohtani H; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Saotome M; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan.
  • Maekawa Y; Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan. Email: ymaekawa@hama-med.ac.jp.
J Invasive Cardiol ; 36(2)2024 Feb.
Article em En | MEDLINE | ID: mdl-38335506
ABSTRACT

OBJECTIVES:

The Impella (Abiomed) is a widely used percutaneous mechanical circulatory support device for high-risk percutaneous coronary intervention in patients with cardiogenic shock. This study aimed to determine the safety and feasibility of a non-angio-guided post-closure approach using the Perclose ProGlide (Abbott) to decannulate the Impella in the intensive care unit.

METHODS:

This retrospective study included consecutive patients who were successfully weaned from mechanical circulatory support using the Impella device between April 2019 and April 2022 at Hamamatsu University School of Medicine. Fifteen patients underwent complete post-closure of the femoral artery access sites at the bedside. Technical success of the post-closure hemostasis technique was defined as no evidence of bleeding or additional medical procedures after manual compression. The safety endpoints comprised the Valve Academic Research Consortium-3 and Bleeding Academic Research Consortium criteria.

RESULTS:

All patients achieved successful hemostasis with this novel technique without surgical conversion. There was no significant bleeding; however, procedure-related vessel occlusion was observed in 1 patient who was recanalized with balloon angioplasty.

CONCLUSIONS:

Bedside post-closure using the Perclose ProGlide device is a safe and feasible alternative to manual compression and surgical removal of the Impella device with low bleeding or vascular complications rates.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Coração Auxiliar / Intervenção Coronária Percutânea Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Invasive Cardiol / J. invasive cardiol / Journal of invasive cardiology Assunto da revista: CARDIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Coração Auxiliar / Intervenção Coronária Percutânea Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Invasive Cardiol / J. invasive cardiol / Journal of invasive cardiology Assunto da revista: CARDIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Japão