Validation of the RBP-9801 oscillometric blood pressure monitor in the general population according to the association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Universal Standard.
Yang, Shijie; Zhou, Zhanyang; Miao, Huanhuan; Zhou, Qiong; Zhai, Mei; Zhang, Yuqing.
Afiliação
  • Yang S; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Zhou Z; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Miao H; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Zhou Q; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Zhai M; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Zhang Y; Department of Cardiology, Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
J Clin Hypertens (Greenwich) ; 26(6): 703-707, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38650108
ABSTRACT
The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor RBP-9801 developed for office and home BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-22018). Subjects were recruited to fulfil the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. A total of 105 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 2.3 ± 6.4/3.1 ± 5.8 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.24/5.03 mmHg (systolic/diastolic). The conclusion is that the RBP-9801 oscillometric device for office and home BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-22018) in the general population and can be recommended for clinic and self-use at home.
Assuntos
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Oscilometria / Determinação da Pressão Arterial / Hipertensão Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Hypertens (Greenwich) Assunto da revista: ANGIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Bases de dados: MEDLINE Assunto principal: Oscilometria / Determinação da Pressão Arterial / Hipertensão Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Hypertens (Greenwich) Assunto da revista: ANGIOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China