Evaluation of lesion contrast in the walk-through long axial FOV PET scanner simulated with XCAT anthropomorphic phantoms.
Dadgar, Meysam; Maebe, Jens; Vandenberghe, Stefaan.
Afiliação
  • Dadgar M; Department of Electronics and Information Systems, Medical Image and Signal Processing, Ghent University, C. Heymanslaan 10, Ghent, Belgium. meysam.dadgar@ugent.be.
  • Maebe J; Department of Electronics and Information Systems, Medical Image and Signal Processing, Ghent University, C. Heymanslaan 10, Ghent, Belgium.
  • Vandenberghe S; Department of Electronics and Information Systems, Medical Image and Signal Processing, Ghent University, C. Heymanslaan 10, Ghent, Belgium.
EJNMMI Phys ; 11(1): 44, 2024 May 09.
Article em En | MEDLINE | ID: mdl-38722428
ABSTRACT

BACKGROUND:

This study evaluates the lesion contrast in a cost-effective long axial field of view (FOV) PET scanner, called the walk-through PET (WT-PET). The scanner consists of two flat detector panels covering the entire torso and head, scanning patients in an upright position for increased throughput. High-resolution, depth-of-interaction capable, monolithic detector technology is used to provide good spatial resolution and enable detection of smaller lesions.

METHODS:

Monte Carlo GATE simulations are used in conjunction with XCAT anthropomorphic phantoms to evaluate lesion contrast in lung, liver and breast for various lesion diameters (10, 7 and 5 mm), activity concentration ratios (81, 41 and 21) and patient BMIs (18-37). Images were reconstructed iteratively with listmode maximum likelihood expectation maximization, and contrast recovery coefficients (CRCs) were obtained for the reconstructed lesions.

RESULTS:

Results shows notable variations in contrast recovery coefficients (CRC) across different lesion sizes and organ locations within the XCAT phantoms. Specifically, our findings reveal that 10 mm lesions consistently exhibit higher CRC compared to 7 mm and 5 mm lesions, with increases of approximately 54% and 330%, respectively, across all investigated organs. Moreover, high contrast recovery is observed in most liver lesions regardless of diameter or activity ratio (average CRC = 42%), as well as in the 10 mm lesions in the lung. Notably, for the 10 mm lesions, the liver demonstrates 42% and 62% higher CRC compared to the lung and breast, respectively. This trend remains consistent across lesion sizes, with the liver consistently exhibiting higher CRC values compared to the lung and breast 7 mm lesions show an increase of 96% and 41%, while 5 mm lesions exhibit approximately 294% and 302% higher CRC compared to the lung and breast, respectively.

CONCLUSION:

A comparison with a conventional pixelated LSO long axial FOV PET shows similar performance, achieved at a reduced cost for the WT-PET due to a reduction in required number of detectors.