Your browser doesn't support javascript.
loading
The influence of undetected hemolysis on POCT potassium results in the emergency department.
Tintu, Andrei N; Buño Soto, Antonio; Van Hoof, Viviane; Bench, Suzanne; Malpass, Anthony; Schilling, Ulf Martin; Rooney, Kevin; Oliver Sáez, Paloma; Relker, Lasse; Luppa, Peter.
Afiliação
  • Tintu AN; Department of Clinical Chemistry Rotterdam, Erasmus Medical Center, Zuid-Holland, Netherlands.
  • Buño Soto A; Clinical Pathology, 16268 Hospital Universitario La Paz , Madrid, Spain.
  • Van Hoof V; Faculty of Medicine and Health Sciences, 26660 University of Antwerp , Wilrijk, Belgium.
  • Bench S; 8945 Guys and St Thomas NHS Trust , London, UK.
  • Malpass A; IDS, Formerly of Becton and Dickinson UK Ltd, Wokingham, Berkshire, UK.
  • Schilling UM; 56750 Linkoping Univ , Linköping, Sweden.
  • Rooney K; 59837 Royal Alexandra Hospital , Paisley, UK.
  • Oliver Sáez P; Laboratory Medicine, 16268 La Paz - Cantoblanco - Carlos III University Hospital , Madrid, Spain.
  • Relker L; Institute for Clinical Chemistry and Pathobiochemistry, 9184 Eberhard Karls Universitat Tubingen , Tubingen, Germany.
  • Luppa P; Institut für Klinische Chemie, 9184 Klinikum rechts der Isar der Technischen Universitat Munchen , Munchen, Germany.
Clin Chem Lab Med ; 2024 May 13.
Article em En | MEDLINE | ID: mdl-38726766
ABSTRACT

OBJECTIVES:

This study aimed to evaluate discrepancies in potassium measurements between point-of-care testing (POCT) and central laboratory (CL) methods, focusing on the impact of hemolysis on these measurements and its impact in the clinical practice in the emergency department (ED).

METHODS:

A retrospective analysis was conducted using data from three European university hospitals Technische Universitat Munchen (Germany), Hospital Universitario La Paz (Spain), and Erasmus University Medical Center (The Netherlands). The study compared POCT potassium measurements in EDs with CL measurements. Data normalization was performed in categories for potassium levels (kalemia) and hemolysis. The severity of discrepancies between POCT and CL potassium measurements was assessed using the reference change value (RCV).

RESULTS:

The study identified significant discrepancies in potassium between POCT and CL methods. In comparing POCT normo- and mild hypokalemia against CL results, differences of -4.20 % and +4.88 % were noted respectively. The largest variance in the CL was a +4.14 % difference in the mild hyperkalemia category. Additionally, the RCV was calculated to quantify the severity of discrepancies between paired potassium measurements from POCT and CL methods. The overall hemolysis characteristics, as defined by the hemolysis gradient, showed considerable variation between the testing sites, significantly affecting the reliability of potassium measurements in POCT.

CONCLUSIONS:

The study highlighted the challenges in achieving consistent potassium measurement results between POCT and CL methods, particularly in the presence of hemolysis. It emphasised the need for integrated hemolysis detection systems in future blood gas analysis devices to minimise discrepancies and ensure accurate POCT results.
Palavras-chave

Texto completo: 1 Bases de dados: MEDLINE Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Bases de dados: MEDLINE Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda